1. Safety 

DANGER!

Failure to comply with these instructions may result in material damage 

or pose a danger to patients or users.

WARNING!

These warnings must absolutely be complied with to guarantee 

safe operation of the product and to avoid any danger to users and to 

patients.

 

NOTE!

Important information: please read carefully. 

1.1  Areas of application of the device

The device is intended to use in professional health care facility environment, like 

doctor's practices, hospitals and optometrists and opticians premises, except near 

of HF surgical equipment and RF shielded rooms of ME‑systems for magnetic res‑

onance imaging. Some portable radio frequency equipment, like cell phones or RF 

telephone equipment including antennas may interference medical devices. Such 

equipment has to be kept in a distance of more than 30 cm (12 inches) from any part 

of the instrument. Inobservance of this precaution may lower the correct function of 

the instrument. The software may be stopped or needs to be restarted. If such unex‑

pected disturbances of the software are observed, the cause could be a cell phone 

or RF telephone in the immediate vicinity to the instrument. Increase the distance 

to the unit, until the interference disappears. The communication between imaging 

module and PC may also be disturbed or interrupted, if the device is exposed to a 

mains power supply delivering excessively transient disturbances or short voltage in‑

terruptions. If this happens, the USB connector needs to be disconnected for a short 

time or the PC needs to be restarted.

1.2  Ambient conditions

Transport:

Temperature

Air pressure

Relative humidity

from

from

from

−40°C

500 hPa

10%

to

to

to

+70°C

1060 hPa

95%

Storage:

Temperature

Air pressure

Relative humidity

from

from

from

−10°C

700 hPa 

10%

to

to

to

+55°C

1060 hPa

95%

Use:

Temperature

Air pressure

Relative humidity

from

from

from

+10°C

800 hPa 

30%

to

to

to

+35°C

1060 hPa

90%

Transport:

Temperature

Air pressure

Relative humidity

from

from

from

−40°C

500 hPa

10%

to

to

to

+70°C

1060 hPa

95%

Storage:

Temperature

Air pressure

Relative humidity

from

from

from

−10°C

700 hPa 

10%

to

to

to

+55°C

1060 hPa

95%

Use:

Temperature

Air pressure

Relative humidity

from

from

from

+10°C

800 hPa 

30%

to

to

to

+35°C

1060 hPa

90%

1.3  Shipment and unpacking

• 

Before you unpack the device, check whether the packaging shows traces of in‑

correct handling or damage. If this is the case, notify the transport company that 

has delivered the goods to you. Unpack the device together with a representative 

of the transport company. Prepare a report for any possible damaged parts. This 

report must be signed by you and by the representative of the transport company.

• 

Leave the device in the packaging for a few hours before unpacking it (conden‑

sation).

• 

Check the device for damage after it is unpacked. Return defective devices in the 

appropriate packaging.

• 

Store packaging material carefully, so that it can be used for possible returns or 

when moving.

1.4  Installation warnings 

WARNING!

• 

Do not modify this equipment without authorization of the manufacturer.

Installation and repairs may only be performed by trained specialists.

Any third-party device must be connected in compliance with the 

EN 60601-1 standard.

Only original Haag-Streit (HS) replacement parts may be used. 

The device must not be stacked or placed in close proximity to other 

electronic devices.

Grounding reliability can only be achieved when unit is connected to 

a hospital grade receptacle. (Not valid for EU countries).

 



 

NOTE!

The power supply unit's mains connector must be accessible in order 

to allow for disconnection from the mains at any time.

1.5  Operation, environment

DANGER!

Never use the device in potentially explosive environments where vola-

tile solvents (alcohol, petrol, etc.) and flammable anaesthetics are in

use.

WARNING!

The device must be switched off after every use. Otherwise there is a

risk of overheating when a protective dust cover is used.

01-IFU_IM600-7220506-04050_eng.indd   4

11.06.2019   16:02:56

Содержание IM 600

Страница 1: ...A HAAG STREIT AG 3098 Koeniz Switzerland HS Doc No 1500 7220506 04050 2019 06 DOC No 1500 1500 1400209 04000 INSTRUCTIONS FOR USE Imaging Module IM 600 Accessory for slit lamps BQ 900 and BP 900 5 edi...

Страница 2: ...urpose of use This device is an accessory for Haag Streit slit lamps BQ 900 or BP 900 which can be used to produce digital photographs and videos for documentation of the eye Contraindication There ar...

Страница 3: ...igger 10 5 4 White balance 10 5 4 1 Slit lamp preparation 10 5 4 2 Conducting a white balance 10 5 5 Software Help menu error messages 11 5 6 LED display illumination head 11 5 7 LED display power sup...

Страница 4: ...hPa 95 Use Temperature Air pressure Relative humidity from from from 10 C 800 hPa 30 to to to 35 C 1060 hPa 90 Air pressure Relative humidity from from 500 hPa 10 to to 1060 hPa 95 Storage Temperature...

Страница 5: ...xpressly disclaims liability for incidental or consequential damage resulting from the use of the product This product is covered by a limited warranty granted by your seller This product is covered b...

Страница 6: ...umination see separate instructions for use 3 Fiber optic line for periphery or background illumination 4 Headrest see separate instructions for use 5 Rail cover 2 4 Camera module CM02 CM02BP 6 Camera...

Страница 7: ...at the corner opposite the black surface Position the sticker against the right roller rail 18 and the gliding plate 19 Press firmly on the white black surface press away any air bubbles Carefully te...

Страница 8: ...connection plug If the middle LED lights up red during operation the two pole connection plug 39 is connected incorrectly Disconnect the device from the pow er grid Remove the cover on the upper part...

Страница 9: ...se module RM02 51 RM02 operational control LED 52 On Off key 53 Rotating knob periphery or back ground illumination 54 Rotating knob slit illumination 55 Selector key A 56 Selector key B 57 Release ke...

Страница 10: ...ease key 57 on the release module RM02 again the image is saved 5 4 White balance The Haag Streit IM 600 is optimized for maximum image quality with the Haag St reit slit lamp BQ 900 and BP 900 The im...

Страница 11: ...illumination Polarity Slit lamp illumination a b c Operating status Standby mode Green short flashes Green short flashes Normal operation Slit and periphery or background illumi nation on Green Green...

Страница 12: ...ached Red flashing Red flashing E4 No communication between power supply and illumination Contact your Haag Streit representative Red flashing 2 Red flashing 2 E6 General error Send PS LED to the appr...

Страница 13: ...M if third party applications such as patient ad ministration software are to be used alongside EyeSuite Use 2 RAM module Hard disk At least 500 GB NTFS data system Graphics Graphics card with at leas...

Страница 14: ...nst liquids A Appendix A 1 Accessories consumables spare parts upgrade Components REF Set white balance greycards 1021485 Power supply ICCNEXERGY 1022106 Computer cable USB 3 0 1022373 Release module...

Страница 15: ...be installed in accordance with the EMC instruc tions contained in this accompanying document The operation of other lines or equipment than those listed may lead to higher emissions or may reduce th...

Страница 16: ...t of a typical commercial or hospital environment Surge EN 61000 4 5 1 kV for symmetrical voltages 1 kV for symmetrical voltages Mains power quality should be that of a typical commercial or hospital...

Страница 17: ...aration distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference...

Страница 18: ...output of the transmitter W Safe distance according to transmission frequency m 150 kHz 80 MHz D 0 7 80 MHz 800 MHz D 1 2 800 MHz 2 5 GHz D 2 3 0 01 0 1 1 10 100 For transmitters with a nominal outpu...

Страница 19: ...EUTSCH ENGLISH FRAN AIS ITALIANO ESPA OL NEDERLANDS PORTUGU S SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc No 1500 7220506 04050 2019 06 01 IFU_IM600 7220506 04050_eng indd 19 11 06 2019 16 0...

Страница 20: ...u have any further questions please contact your Haag Streit representative at http www haag streit com contact contact your distributor html HAAG STREIT AG Gartenstadtstrasse 10 3098 Koeniz Switzerla...

Отзывы: