F.4  Immunity for non-life support devices (standard table 4)

The information is based on the requirements of EN 60601‑1‑2:2007 (IEC 3

rd

 edition).

Guidance and manufacturer's declaration – electromagnetic immunity

This product is intended for use in the electromagnetic environment specifi ed below. The customer or the user of this product should assure that it is used in such an envi-

ronment.

Electromagnetic environment – guidance

Portable and mobile RF communications equipments hould be used no closer to any part of this product, including cables, than the recommended separation distance cal-

culated from the equation applicable to the frequency of the transmitter.

Immunity test standard

EN 60601 test level

Compliance level

Recommended distance

(c)

:

Conducted RF EN 61000-4-6

3 V

rms

 

150 kHz – 80 MHz

5 V

rms

D

 = 0.7 

Radiated RF EN 61000-4-3

3 V/m

80 MHz – 2.7 GHz

3 V/m

80 MHz – 2.7 GHz

D

 = 1.2   80 MHz – 800 MHz

D

 = 2.3   800 MHz – 2.7 GHz

Where 

P

 is the maximum output power rating of thet ransmitter in watts 

(W)

 according to the transmitter manufacturer and 

D

 is the recommended separation 

distance in metres 

(m)

. Field strengths from fi xed RF transmitters, as determined by an electromagnetic site survey

 a

, should be less than the compliance level 

in each frequency range

 b

 Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1: At 80 MHz and 800 MHz the higher frequency applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

 

a.  Field strengths from fi xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broad-

cast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fi xed RF transmitters, an electromagnetic 

site survey should be considered. If the measured fi eld strength in the location in which this product is used exceeds the applicable RF compliance level above, this 

product should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating 

this product.

b.

c. 

Over the frequency range 150 kHz to 80 MHz, 

Possible shorter distances outside the ISM bands do not contribute to improved application in this table.

fi eld strengths should be less than 5 V

eff

.

01-IFU_IM600-7220506-04050_eng.indd   17

11.06.2019   16:03:08

Содержание IM 600

Страница 1: ...A HAAG STREIT AG 3098 Koeniz Switzerland HS Doc No 1500 7220506 04050 2019 06 DOC No 1500 1500 1400209 04000 INSTRUCTIONS FOR USE Imaging Module IM 600 Accessory for slit lamps BQ 900 and BP 900 5 edi...

Страница 2: ...urpose of use This device is an accessory for Haag Streit slit lamps BQ 900 or BP 900 which can be used to produce digital photographs and videos for documentation of the eye Contraindication There ar...

Страница 3: ...igger 10 5 4 White balance 10 5 4 1 Slit lamp preparation 10 5 4 2 Conducting a white balance 10 5 5 Software Help menu error messages 11 5 6 LED display illumination head 11 5 7 LED display power sup...

Страница 4: ...hPa 95 Use Temperature Air pressure Relative humidity from from from 10 C 800 hPa 30 to to to 35 C 1060 hPa 90 Air pressure Relative humidity from from 500 hPa 10 to to 1060 hPa 95 Storage Temperature...

Страница 5: ...xpressly disclaims liability for incidental or consequential damage resulting from the use of the product This product is covered by a limited warranty granted by your seller This product is covered b...

Страница 6: ...umination see separate instructions for use 3 Fiber optic line for periphery or background illumination 4 Headrest see separate instructions for use 5 Rail cover 2 4 Camera module CM02 CM02BP 6 Camera...

Страница 7: ...at the corner opposite the black surface Position the sticker against the right roller rail 18 and the gliding plate 19 Press firmly on the white black surface press away any air bubbles Carefully te...

Страница 8: ...connection plug If the middle LED lights up red during operation the two pole connection plug 39 is connected incorrectly Disconnect the device from the pow er grid Remove the cover on the upper part...

Страница 9: ...se module RM02 51 RM02 operational control LED 52 On Off key 53 Rotating knob periphery or back ground illumination 54 Rotating knob slit illumination 55 Selector key A 56 Selector key B 57 Release ke...

Страница 10: ...ease key 57 on the release module RM02 again the image is saved 5 4 White balance The Haag Streit IM 600 is optimized for maximum image quality with the Haag St reit slit lamp BQ 900 and BP 900 The im...

Страница 11: ...illumination Polarity Slit lamp illumination a b c Operating status Standby mode Green short flashes Green short flashes Normal operation Slit and periphery or background illumi nation on Green Green...

Страница 12: ...ached Red flashing Red flashing E4 No communication between power supply and illumination Contact your Haag Streit representative Red flashing 2 Red flashing 2 E6 General error Send PS LED to the appr...

Страница 13: ...M if third party applications such as patient ad ministration software are to be used alongside EyeSuite Use 2 RAM module Hard disk At least 500 GB NTFS data system Graphics Graphics card with at leas...

Страница 14: ...nst liquids A Appendix A 1 Accessories consumables spare parts upgrade Components REF Set white balance greycards 1021485 Power supply ICCNEXERGY 1022106 Computer cable USB 3 0 1022373 Release module...

Страница 15: ...be installed in accordance with the EMC instruc tions contained in this accompanying document The operation of other lines or equipment than those listed may lead to higher emissions or may reduce th...

Страница 16: ...t of a typical commercial or hospital environment Surge EN 61000 4 5 1 kV for symmetrical voltages 1 kV for symmetrical voltages Mains power quality should be that of a typical commercial or hospital...

Страница 17: ...aration distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference...

Страница 18: ...output of the transmitter W Safe distance according to transmission frequency m 150 kHz 80 MHz D 0 7 80 MHz 800 MHz D 1 2 800 MHz 2 5 GHz D 2 3 0 01 0 1 1 10 100 For transmitters with a nominal outpu...

Страница 19: ...EUTSCH ENGLISH FRAN AIS ITALIANO ESPA OL NEDERLANDS PORTUGU S SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc No 1500 7220506 04050 2019 06 01 IFU_IM600 7220506 04050_eng indd 19 11 06 2019 16 0...

Страница 20: ...u have any further questions please contact your Haag Streit representative at http www haag streit com contact contact your distributor html HAAG STREIT AG Gartenstadtstrasse 10 3098 Koeniz Switzerla...

Отзывы: