| SERVICE MANUAL / INSTALLATION GUIDE
6
© HAAG-STREIT AG, 3098 Koeniz, Switzerland
–
HS-Doc. No. 7220676 - Edition 2.00 / 2020
2.
Overview
3.
Installation
NOTE!
The software and an installation manual are included in the scope of delivery.
Please install the latest EyeSuite Software first and then the FUNDUS MODULE 300 software.
WARNING!
Installation and repairs may only be performed by authorized specialists. Contact your HAAG-
STREIT representative for installation, repairs and modification work on the system. The contact
details are available at
WARNING!
•
No modification of this equipment is allowed.
•
It is only allowed to attach USB cables and Power source provided in the sales package. If you
need replacement to the USB cables or power source, please contact your local dealer.
•
USB cable must be connected only to the USB port of a PC that complies with the IEC 60950
standard.
•
Only connect the FUNDUS MODULE 300 and Release Module RM02 to an electricity supply
which corresponds to the supply voltage of the devices.
•
Use only accessories and battery provided by HAAG-STREIT with this product.
•
When connecting the device, the operator is responsible for operating of the device and for
meeting the safety and EMC requirements defined in IEC 60601-1-1 or IEC 60601-1:2005,
chapter 16 (medical electrical systems) and IEC 60601-1-2.
•
Additional extension cables or portable multiple sockets may not be used.
•
The electrical installation must conform to IEC 60364-7-710 (for USA and CANADA only: for
single-phase power systems with 120 V AC, NEMA 5-15P power plugs must be used) or the
applicable national regulations. This includes the existence of a ground fault circuit interrupter
(GFCI).
•
To avert the risk of an electric shock this device may only be connected to a power supply with
a protective earth conductor.
•
The device should be set up so that the power cable can be disconnected from the power
supply quickly and easily without any tools.
•
The cables for the accessories should be laid to prevent any risk of the examiner tripping over
them or their presenting any danger to the patient.
Fundus Module 300 Camera
Lens cover
Fundus Module 300 Cradle
Battery
Power Supply
Cradle USB cable
Base plate
Camera USB cable