| SERVICE MANUAL / INSTALLATION GUIDE
© HAAG-STREIT AG, 3098 Koeniz, Switzerland
–
HS-Doc. No. 7220676 - Edition 2.00 / 2020
15
3.8.2
FUNDUS MODULE 300 connected to medical PC
Fig. 26
A.
Power grid
B.
Medically approved isolating transformer
C.
Instrument table (IT)
D.
Medically approved power supply
E.
Printer
F.
Screen
G.
Personal computer
H.
Local network
I.
Galvanic isolation (EN 60601-1)
J.
Headrest
K.
Cable headrest/LED illumination/RM02
L.
Release module RM02
M.
LED illumination
N.
FUNDUS MODULE 300 Camera
O.
FUNDUS MODULE 300 Cradle
P.
Imaging Module (IM 600/IM 900/Imaging set for BI 900)
3.9
Step-by-step cabling
WARNING!
The Release Module RM02 is affixed with strong magnets. Keep magnet-sensitive storage media (e.g.
credit card) away from the magnets.
Computer cable USB 3.0 (a), power supply cable (b)
and cable headrest/LED illumination/RM02 (c) must
be fed through in the braided sleeving.
Fig. 27
a
b
c