Please follow national and local infection control laws, standards or guidelines.
Please observe the manufacturer’s instructions for all cleaning and disinfecting
agents.
The manufacturer, Haag-Streit UK, only approves the following cleaning and
disinfecting procedure. Other procedures and cleaning agents must be
validated.
Only manufacturer-recommended bulbs should be used*. Please verify the lamp
voltage complies with the supply voltage.
To replace the G4 6v 10W halogen bulb, remove the protective bulb cover using the tool
supplied (see Figure 15). Rotate a quarter turn, counter clockwise. Engage the bulb and
pull it outwards. Insert the replacement bulb into the tool, ensuring the pins are
horizontal. Insert the pins into the lamp housing and release the tool by pressing the
plunger (see Figure 16). Replace the protective bulb cover, reconnect the power supply
and test the slit lamp.
Figure 15
Figure 16
7.4 Cleaning & disinfection
REMOVING THE BULB COVER
PRESS THE PLUNGER
Do not touch the glass body of the bulb.
*For details of manufacturer-recommended bulbs, please see appendix A.