Haag-Streit AG

Operation of C.MOR HD and C.MOR HD³ | 7

1500 4000089 02010_web

23

Button

Function

Create and Save Still Images

Open and Select Stored Scene Files.

7.2 Switch the CCU on and off

The camera head is automatically switched on or off with the surgical microscope system. 
To switch the CCU on or off:

1. Before turning on the power of the CCU, make sure that all devices (camera head, monitors)

are connected correctly.

2. Press the power button.

ð

If the power button lights up green, the video camera is switched on.

ð

If the power button is not lit, the video camera is switched off.

7.3 Configure Setup Menu

This chapter only shows the most frequently required menu settings that can be used to improve
the camera image under a wide range of environmental conditions.

Different types of surgery require different types of illumination. The illumination of the surgical field
can be influenced by using different light sources (e.g. endoilluminator, microscope light), different
temperatures of light (e.g. halogen filter, daylight filter) or the light intensity.

In addition, the camera image can look different on an external monitor, so that the camera image
differs from the microscope’s image.

The camera image can be adjusted according to individual preferences, which affects the quality of
the image.

An overview of all possible menu settings and their default settings are shown in chapter 

Menu Set-

tings [

}

 24]

.

Содержание C.MOR HD

Страница 1: ...Surgical Microscope Cameras C MOR HD C MOR HD Instructions for Use...

Страница 2: ...nd Disinfection 18 6 2 1 Cleaning 18 6 2 2 Disinfection 20 6 2 3 Visual Guide for Cleaning and Disinfection 21 7 Operation of C MOR HD and C MOR HD 22 7 1 Symbols for Control Elements 22 7 2 Switch th...

Страница 3: ...1500 4000089 02010_web iii 9 1 General 35 9 2 Maintenance Intervals 35 10 Disposable Materials and Spare Parts 36 11 Technical Data 37 12 Disposal 39 13 Appendix 40 13 1 Specifications on Electromagne...

Страница 4: ...e from the respective components type plates You will find the contact details for your local sales representative in the notes pages of these in structions A current list of local sales representativ...

Страница 5: ...Haag Streit AG Manufacturer s Details and Contact Information 1 1500 4000089 02010_web 5 Item Name 12 Product Name...

Страница 6: ...or death Follow the instructions to avoid a hazard CAUTION May lead to light or moderately serious injuries Follow the instructions to avoid a hazard NOTICE May lead to material damage and or operati...

Страница 7: ...of limited treatment options If defects are not detected the surgical microscope system may not offer all functions during sur gery a Before each use make sure that all electrical and mechanical conne...

Страница 8: ...tate in which the user and or the patient is established First Installation and Commissioning The first installation and commissioning of the medical device must be done in accordance with the instruc...

Страница 9: ...from third party suppliers have to be installed After maintenance or technical modifications the device must be readjusted following the service manual of the surgical microscope system In case of te...

Страница 10: ...5 2 1 For connecting the power supply unit to the CCU 3 Power Supply Unit Supplies the CCU with power 4 IEC Socket C13 For connecting the power supply unit to the mains power socket 5 M23 Round Socket...

Страница 11: ...ead Reference number REF serial number SN Health In dustry Bar Code HIBC of the camera head 2 CCU Connection For connecting the camera head to the camera control unit CCU 3 C Mount Connection For conn...

Страница 12: ...tting the camera functions see Symbols for Control Elements 22 2 Menu Navigation Controls Buttons for selecting menu items and determining settings see Symbols for Control Elements 22 3 Button FUNCTIO...

Страница 13: ...camera cable 6 IRIS For connecting an automatic aperture lens Not intended for use with the surgical microscope system 7 Power For connecting the power supply unit 8 SYNC Only for the 3D view functio...

Страница 14: ...he camera head via camera cable 6 Power For connecting the power supply unit 7 SYNC Output for synchronizing when using two sets of cameras and CCUs Sends synchronization signals to the slave CCU 8 GE...

Страница 15: ...rea can injure the patient a Ensure careful attachment of accessories 5 1 Connect the Camera Head For attaching the camera head a camera attachment with C mount adapter must be installed to the micros...

Страница 16: ...the CCU onto the tray of the carrier unit 2 Connect the monitor to the CCU For using the VBS signal analog SD quality connect a BNC cable to the VBS output socket For using the S Video signal analog S...

Страница 17: ...r supply unit into the mains power socket 5 3 Check the Image Alignment 1 Make sure that the orientation of the camera image matches the orientation of the micro scope s image The left and right sides...

Страница 18: ...provided by the manufacturer of the disinfectant regarding recommended concentration and contact time After all appropriate items are cleaned disinfected or sterilized they should be inspected 1 Chec...

Страница 19: ...t Let disinfectant dry out in air External surfaces Materials required Disposable gloves Cloth Damp cloth single use tissues or wipes cotton or mi crofiber cloth Cleaning agent glutaraldehyde based or...

Страница 20: ...to surfaces 3 Use slightly soaked wipes with a sufficient amount of disinfectant 4 Let disinfected surfaces dry out in air 5 Be aware of the exposure time with the disinfectant 6 Be aware of the appli...

Страница 21: ...ning and Disinfection Color Note Green Parts that are intended for cleaning and disinfection Black Parts that are intended to be touched by the user They must be cleaned and disinfected before and aft...

Страница 22: ...on on HD MIOS or the external monitor works b In case of poor image quality reset the CCU to the default settings see Menu Settings 24 7 1 Symbols for Control Elements Button Function Switch the CCU o...

Страница 23: ...enu This chapter only shows the most frequently required menu settings that can be used to improve the camera image under a wide range of environmental conditions Different types of surgery require di...

Страница 24: ...ame function as hue adjustment via the button PAINT AE See Adjust Hue and Automatic Exposure PAINT AE 32 Gain Offset Blue 0 Adjusts the blue level Same function as hue adjustment via the button PAINT...

Страница 25: ...Gain Control is set to AUTO Area Select NARROW Adjusts the extent of the measuring area for the automatic sensitivity settings based on the image that is displayed on the mon itor see Adjust Automati...

Страница 26: ...3D Support SLAVE Sets the function of the CCU when 3D support is used Picture Setting DNR ON Activates the noise reduction DNR Level 8 Adjusts the noise reduction strength In creasing the value reduc...

Страница 27: ...cene files Only for C MOR HD 7 3 2 Show or Hide Setup Menu 1 Make sure that an external monitor is connected to the CCU see Installation to the Surgical Microscope 15 2 Press the button MENU on the fr...

Страница 28: ...at to the desired value 59 94 Hz The VBS signal outputs the NTSC format with a frame rate of 59 94 frames per second 50 Hz The VBS signal outputs the PAL format with a frame rate of 50 frames per seco...

Страница 29: ...continuously and thereby white balance is set automatically This function is recom mended when lighting conditions change frequently 6 Confirm the entry by pressing the button FUNCTION SET on the fro...

Страница 30: ...To change settings in the setup menu the menu lock must be deactivated 1 Open the Setup Menu 2 Select the menu item Miscellaneous 3 Set the value of the menu item Menu Lock to OFF The red lock mark in...

Страница 31: ...item Menu Lock to ON 7 4 Perform Automatic White Balance AWB Whenever the lighting conditions change the white balance should be performed Otherwise the color reproduction of the camera image will be...

Страница 32: ...temporary and will be reset if the CCU is switched off 1 Press the button PAINT AE on the front panel of the CCU to activate the hue adjustment The button PAINT AE lights up green and the hue adjustme...

Страница 33: ...lor bar test pat tern If the button BARS lights up green the color bar test pattern is shown If the button BARS is not lit the color bar test pattern is hidden 7 8 Create Freeze Images FREEZE REC C MO...

Страница 34: ...vice NOTICE Resettable error reoccurs twice or more The error may be due to a systematic fault in the medical device and must be attended by Haag Streit Please contact your local representative or Haa...

Страница 35: ...ot described in these instructions must be conducted by authorized service technicians Maintenance tasks which can be conducted by the user are described in the following table User Maintenance Interv...

Страница 36: ...e manufacturer as spare parts for internal components may increase electromagnetic emissions or reduce the system s immunity to interference a Only use approved accessories and spare parts b Do not ex...

Страница 37: ...ance class I equipment EMC IEC 60 601 1 2 UMDNS 12 539 GMDN 12539 Degree of protection by enclosure IP Code IP20 Technical Data C MOR HD C MOR HD REF 4000011 4000012 Mechanical data Dimensions CCU Hei...

Страница 38: ...resolution 900 TV lines 1000 TV lines Connections VBS PAL NTSC PAL NTSC Y C S Video S Video RS 232C not used not used FOOT SW not used not used Camera Connection of the camera head Connection of the c...

Страница 39: ...European Union the surgical microscope system and its components are subject to EU Directive on Waste of Electrical and Electronic Equipment and may not be disposed together with waste from private ho...

Страница 40: ...0 MHz 9 V m 704 MHz 787 MHz 28 V m 800 MHz 960 MHz 28 V m 1 7 GHz 1 99 GHz 28 V m 2 4 GHz 2 57 GHz 9 V m 5 1 GHz 5 8 GHz Electrical fast transient disturb ance bursts IEC 61000 4 4 For signal lines 1...

Страница 41: ...HD SDI The HD SDI is a Serial Digital Interface primar ily for the transmission of uncompressed and un encrypted High Definition video data via coaxial cable HIBC HIBC is the abbreviation of the term...

Страница 42: ...of the video picture signal V the blank ing signal B and the synchronization signal S The VBS signal is an analog video signal This transmission technique is also called composite video Y C Y C stands...

Страница 43: ...Haag Streit AG 1500 4000089 02010_web 43...

Страница 44: ...HS Art No 1500 4000089 02010_web 1 Edition 2024 02 Haag Streit AG Gartenstadtstrasse 10 3098 Koeniz Switzerland Phone 41 31 978 01 11 mail info haag streit com web www haag streit com...

Отзывы: