Operation of C.MOR HD and C.MOR HD³ | 7
1500 4000089 02010_web
31
Store a scene file
1. Make sure that all settings that are to be saved in this scene file are correctly adjusted.
2. Select the menu item
Scene File
.
3. Select
Store Scene
.
4. Select the number of the scene file which should be saved.
5. Confirm the entry by pressing the button [FUNCTION/SET] on the front panel of the CCU.
ð
All current setting values are saved to the scene file.
Load a scene file
To load a stored scene file, see
Initialize a scene file
1. Select the menu item
Scene File
.
2. Select the
Initialize Scene
.
3. Select the number of the scene file which should be initialized or select
All
to initialize all
scene files.
4. Confirm the entry by pressing the button [FUNCTION/SET] on the front panel of the CCU.
5. To activate the menu lock, set the value of the menu item
Menu Lock
to
ON
.
7.4 Perform Automatic White Balance (AWB)
Whenever the lighting conditions change, the white balance should be performed. Otherwise, the
color reproduction of the camera image will be incorrect.
NOTICE
If the position of the microscope’s filter is wrong, the automatic white balance pro-
duces a camera image that is wrong in terms of color.
a) Make sure not to use a filter during the automatic white balance performance.
1. Place a white object under the surgical microscope’s focus module, e.g., a blank piece of
white copy paper.
2. Switch on the illumination that should be used, e.g., the endoilluminator.
3. Press the button [AWB] on the front panel of the CCU.
ð
The button [AWB] lights up green and the operation is displayed on the monitor:
–
AWB BUSY
: The camera is currently performing the automatic white balance.
–
AWB COMPLETED
: The automatic white balance was performed successfully.
–
AWB ERROR
: The automatic white balance was not performed because the illumination is
too bright or too dark. Adjust the illumination and perform the automatic white balance
again.