HY-ALERTA
TM
500 HANDHELD HYDROGEN LEAK
DETECTOR OPERATING MANUAL
90000002 R13
Page 12 of 14
Symptom
Possible Cause
Action
Sensor Error
The probe is disconnected from
the controller.
Turn off the instrument and verify that the
probe is properly connected to the controller.
Error 88
Faulty capacitor
Cycle of power
Error 40
The sensor PCB temperature is
too high.
Turn off the instrument. Let it cool.
Error 20
The sensor temperature is
incorrect.
Cycle of power
Unit won’t turn on
The battery is dead.
Charge the battery completely.
If the recommended action does not solve the problem, the HY-ALERTA 500 should be returned to the
factory for repair.