background image

 

 

 HY-ALERTA

TM

 500 HANDHELD HYDROGEN LEAK 

DETECTOR OPERATING MANUAL 

90000002 R13 

 

Page 4 of 14 

IMPORTANT NOTICES 

 

Read and understand this operating manual before installing or using the unit.   
Only use cables, battery pack, battery charger, and AC/DC power supply from H2scan with this unit. 
 If this equipment is used in a manner not specified by H2scan, the protection provided by this equipment may be 
impaired. 

 

Hydrogen is flammable at 4% in air. Take indications seriously and be prepared to take action.  In the event of 
detection of 4% or higher of a hydrogen gas concentration there is a high probability of a hazard to safety.  Inform 
local emergency response personnel immediately. 

 

Any exposure to condensation will damage the sensor. The sensor die is made of a Palladium Nickel film and is 
sensitive to any form of condensation (organic and aqueous).  Keep the sensor die dry. 

 

Limitation  of  Liability:

  SELLER  SHALL  UNDER  NO  CIRCUMSTANCES  BE  LIABLE  FOR  ANY  INCIDENTAL,  CONSEQUENTIAL,  SPECIAL, 

PUNITIVE, OR OTHER DAMAGES, INCLUDING, BUT NOT LIMITED TO, LOSS OF BUSINESS OR PROFIT, PROMOTIONAL OR MANUFACTURING 
EXPENSES, INJURY TO REPUTATION, OR LOSS OF CUSTOMER, BASED ON ANY ALLEGED NEGLIGENCE, BREACH OF WARRANTY, STRICT 
LIABILITY,  BREACH  OF  CONTRACT,  OR  ANY  OTHER  LEGAL  THEORY  ARISING  OUT  OF  THE  USE,  MISUSE,  PURCHASE,  SALE  OR 
POSSESSION  OF  ITS  GOODS  OR  ITS  PERFORMANCE  OF  THIS  CONTRACT  TO  THE  EXTENT  THAT  SUCH  LIABILITY  EXTEND

S  SELLER’S 

OBLIGATIONS BEYOND THE PRICE PAID BY BUYER TO SELLER FOR THE ITEM ON WHICH SUCH CLAIM IS BASED. SELLER ADVISES BUYER 
TO PERFORM ACCEPTABLE TESTS ON ALL HARDWARE PRIOR TO DEPLOYMENT AND TO PERFORM MAINTENANCE AS DESCRIBED IN THE 
SELLER’S INSTRUCTION GUIDE. UNDER NO CIRCUMSTANCES SHALL THE  EQUIPMENT PROVIDED HEREUNDER BE USED IN  A MANNER 
WHERE  IT  IS  THE  SOLE  PROTECTIVE  SYSTEM  FOR  FACILITIES,  EQUIPMENT  AND  PERSONNEL  SAFETY;  THE  EQUIPMENT  IS  INTENDED 
FOR USE IN CONUNCTION WITH OTHER APPROPRIATE PROTECTIVE SYSTEMS.

 

LIMITED WARRANTY 

H2scan Limited Warranty:

 

Each hydrogen instrument (“Product”) will conform, as to all substantial operational features, to the Product 

specifications set forth this Manual and will be free of defects which substantially affect suc

h Product’s performance for twelve (12) months from 

the ship date for such Product. 

Must Provide Notice of Defect:

 If you believe a Product that you believe is defective, you must notify H2scan in writing, within ten (10) days of 

receipt of such Product, of your claim regarding any such defect. 

Return Product to H2scan for Repair, Replacement or Credit:

 

If the Product is found defective by H2scan, H2scan’s sole obligation under this 

warranty is to either (i) repair the Product, (ii) replace the Product, or (iii) issue a credit for the purchase price for such Product, the particular 
remedy to be determined [by H2scan] on a case-by-case basis. 

Voided Warranty: 

H2scan’s 12 Month Limited Warranty is void for any of the following: 

The unit is opened and the manufacturing seal is broken 

Unauthorized repair work performed at the customer’s location or carried out by anyone other than H2scan’s factory trained technicians 

Equipment or parts that have been tampered with, misused, neglected, mishandled, improperly adjusted, or modified in any way without the 
written consent of H2scan. 

Equipment or parts that have been damaged due to shipping, misuse, accidents, mishandling, neglect, or problems with electrical power sources. 

Repair work performed during the warranty period does not prolong the warranty period past the original period. 

System operation in incorrect or inappropriate environments. 

Usage that is not in accordance with system guidelines or an operator’s failure to follow manual instructions. 

LIMITATION OF WARRANTY:

  THE ABOVE IS A LIMITED WARRANTY AS IT IS THE ONLY WARRANTY MADE 

BY H2SCAN.  H2SCAN MAKES NO OTHER WARRANTY EXPRESS OR IMPLIED AND EXPRESSLY EXCLUDES 
ALL WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.  YOUR SOLE 
REMEDY HEREUNDER IS REPAIR OR REPLACEMENT OF THE PRODUCT OR A CREDIT FOR THE 
PURCHASE PRICE FOR SUCH PRODUCT, THE PARTICULAR REMEDY TO BE DETERMINED BY H2SCAN ON 
A CASE-BY-CASE BASIS.  H2SCAN SHALL HAVE NO LIABILITY WITH RESPECT TO ITS OBLIGATIONS UNDER 
THIS AGREEMENT FOR CONSEQUENTIAL, EXEMPLARY, OR INCIDENTAL DAMAGES EVEN IF IT HAS BEEN 
ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.  THE STATED EXPRESS WARRANTY IS IN LIEU OF ALL 
LIABILITIES OR OBLIGATIONS OF H2SCAN FOR DAMAGES ARISING OUT OF OR IN CONNECTION WITH THE 
DELIVERY, USE OR PERFORMANCE OF THE PRODUCTS. 

Содержание HY-ALERTA 500

Страница 1: ...HY ALERTA 500 Handheld Hydrogen Leak Detector ...

Страница 2: ...ed assets Grid Petroleum refinery and other industrial process control Facility and equipment safety to minimize downtime at a much lower total costs of ownership than the competition Our Strategic Objectives H2scan s technology accepted as the new gold standard in hydrogen sensors H2scan s business recognized for innovation and ingenuity high quality products and systems application specific solu...

Страница 3: ...set Operation 9 4 4 Zero Operation 9 5 KEYPAD 9 5 1 Numerical Changes 9 5 2 Top Level Keypad Functions 9 5 3 General Keypad Functions 10 5 4 Information Display 10 5 5 Firmware Revision 10 5 6 Serial Number 10 5 7 Factory Calibration Date 10 5 8 Reset Sensor 10 5 9 Zero Sensor 10 6 HYDROGEN SENSING CONSIDERATIONS 10 7 BUMP TEST 11 8 BATTERY CHARGING 11 9 CLEANING 11 10 TROUBLESHOOTING 11 11 MENU S...

Страница 4: ...fects which substantially affect such Product s performance for twelve 12 months from the ship date for such Product Must Provide Notice of Defect If you believe a Product that you believe is defective you must notify H2scan in writing within ten 10 days of receipt of such Product of your claim regarding any such defect Return Product to H2scan for Repair Replacement or Credit If the Product is fo...

Страница 5: ...ppm to 100 hydrogen by volume in air Response Time Indication of hydrogen within seconds Stabilization to final value depends on concentration Ambient Temperatures Operating 0 C to 40 C Storage 20 C to 45 C Relative Humidity 0 95 non condensing Power Internal rechargeable Lithium Ion battery yields over 10 hours of operation and is recharged in 4 hours with included charger Battery charger input 1...

Страница 6: ...HY ALERTATM 500 HANDHELD HYDROGEN LEAK DETECTOR OPERATING MANUAL 90000002 R13 Page 6 of 14 Coiled Cord Sensor Probe Controller Controller Carry Pouch Battery Charger with Regional Plugs ...

Страница 7: ...es of operation remaining A fully charged battery should last 10 to15 hours depending on use There is a small load on the battery when the unit is powered off This load will discharge the battery of the unit in it s powered off state in about 6 months Customers that do not use the device frequently should charge the battery and perform a bump test with hydrogen gas every one to three months to kee...

Страница 8: ...graph array indicates detected hydrogen concentrations in four ranges as noted below 0 002 0 005 0 010 0 022 0 046 0 100 0 220 0 460 1 000 2 200 4 600 10 00 22 00 46 00 100 0 0 001 Hydrogen 4 67 Solid Green Unit ready 15 ppm hydrogen Pulsing Green Decreasing H2 level Solid Amber System startup Steady H2 level Pulsing Red Increasing H2 level Alternating Red Green Fluctuating H2 level Probe Tip Colo...

Страница 9: ... THE LCD INDICATING LESS THAN 0 1 HYDROGEN BEFORE INVOKING THE RESET OPERATION 4 4 ZERO OPERATION The Zero Operation is used to zero the hydrogen reading if the instrument is reporting low levels 0 001 to 0 01 of hydrogen when no hydrogen is present It can be invoked from the keypad while in the top menu level measuring hydrogen by pressing and holding left arrow button or from the Zero Menu Zero ...

Страница 10: ...number preceded by a C is the Controller firmware revision For example S1 23 C2 34 for sensor probe firmware version 1 23 and Controller firmware version 2 34 5 6 SERIAL NUMBER This displays the product serial number For example 50123 5 7 FACTORY CALIBRATION DATE This displays the date of last factory calibration MM DD YY For example 5 8 06 for 8 May 2006 5 8 RESET SENSOR The Reset Sensor menu is ...

Страница 11: ...or a leak in areas where liquid hydrogen is stored check both above and below the area of concern 7 BUMP TEST A bump test is recommended every three months The purpose of a bump test is to verify that the sensor is active detecting hydrogen and verifying that the sensor is within the pre set factory tolerance for accuracy To perform a bump test perform the following In a hydrogen free environment ...

Страница 12: ...ature is too high Turn off the instrument Let it cool Error 20 The sensor temperature is incorrect Cycle of power Unit won t turn on The battery is dead Charge the battery completely If the recommended action does not solve the problem the HY ALERTA 500 should be returned to the factory for repair 11 MENU STRUCTURE Firmware Rev Serial Number Factory Calibration date Information Display Reset Senso...

Страница 13: ...MANUAL 90000002 R13 Page 13 of 14 Zero Sensor In Progress Zero Sensor Done ENTER Zero Sensor Continue Reset Sensor Not Done Reset Sensor Done No Reset Sensor Continue Reset Sensor In Progress ENTER Zero Sensor Not Done No Hydrogen Hydrogen Zero Menu Reset Menu ...

Страница 14: ...HY ALERTATM 500 HANDHELD HYDROGEN LEAK DETECTOR OPERATING MANUAL 90000002 R13 Page 14 of 14 12 EUROPEAN DECLARATION OF CONFORMITY ...