![GW Instek GDS-1000B Series Скачать руководство пользователя страница 205](http://html1.mh-extra.com/html/gw-instek/gds-1000b-series/gds-1000b-series_programming-manual_2280038205.webp)
COMMAND DETAILS
205
Note
This setting only applies when the UART search
trigger is set to trigger on Tx Data
Syntax
:SEARCH:TRIGger:BUS:B1:UART:TX:DATa:SIZe
{<NR1> | ?}
Related
commands
:SEARCH:TRIGger:BUS:B1:UART:CONDition
Parameter
<NR1>
Number of bytes (1 to 10).
Return parameter <NR1>
Returns the number of bytes.
Example
:SEARCH:TRIGger:BUS:B1:UART:TX:DATa:SIZe 5
Sets the number of bytes to 5.
:SEARCH:TRIGger:BUS:B1:UART:TX:DATa:
VALue
Description
Sets or queries the search triggering data value for
the UART bus when the bus is set to trigger on Tx
Data.
Syntax
:SEARCH:TRIGger:BUS:B1:UART:TX:DATa:VALue
{<string> | ? }
Related
commands
:SEARCH:TRIGger:BUS:B1:UART:TX:DATa:SIZe
Parameter
<sting>
The number of characters in the string
depends on the data size setting. The
string must be enclosed in double
quotes, "string".
x = don’t care
1 = binary 1
0 = binary 0
Return Parameter
Returns the data value.
Set
Query