OPERATION
81
Buzzer
It enables or disables buzzer sound for the
select step.
Buzzer
ON, OFF
Measure
Average
It sets the speed level of display sampling for
the measure average setting for the select step.
More the average numbers (High), slower the
display update. Refer to page 112 for details.
Measure
Average
Off, Low, Middle, High
Jump To
It sets the target step to jump to. For example,
when step 5 is set for Jump To under the step 2
Edit page, it means that when sequence runs to
step 2, it will directly jump to step 5 at the end
of step 2. The available step option(s) depends
on the total step setting.
Jump To
1 ~ 20000
Jump Count
It sets the number of times to loop the Jump To
step action.
Jump Count
1 ~ 10000
Note
Jump and Cycle functions can Not be activated at
the same time. Refer to page 75 for details of Cycle.
Trigger Out
It sets if trigger out signal will be transmitted
when the sequence runs to the step.
Trigger Out
ON, OFF
Note
When there is any issue occurred from settings,
PPX series will not be able to run sequence script.
The error code along with warning message will be
shown within the prompt message box when Run
filed is enabled.
Содержание 4713008674147
Страница 70: ...PPX Series User Manual 70 Red The alarm of short circuit occurs from temperature measurement...
Страница 151: ...MENU CONFIGURATION 151 Communicati on Monitor Display The icon indicates Communication Monitor is activated...
Страница 172: ...PPX Series User Manual 172 Schematic PPX Trigger input signal Analog connector 12 10 B COM B COM 100k...
Страница 218: ...PPX Series User Manual 218 PPX Dimensions PPX Series...