ANALOG CONTROL
161
Connection
When connecting the external voltage source to
the analog connector, use shielded or twisted
paired wiring.
PPX
EXT-V
Analog
connector
+
-
4
3
Output
Terminal
2 core shielded
wire or twisted
pair
Pin3 → EXT-V (+)
Pin4 → EXT-V (-)
Wire shield → negative (-) output terminal
Connection- alt.
shielding
If the wire shield needs to be grounded at the
voltage source (EXT-V), then the shield cannot
also be grounded at the negative (-) terminal
output of the PPX power supply. This would
short the output.
PPX
EXT-V
Analog
connector
+
-
4
3
Output
Terminal
2 core shielded
wire or twisted
pair
Pin3 → EXT-V (+)
Pin4 → EXT-V (-)
Wire shield → EXT-V ground (GND)
Содержание 4713008674147
Страница 70: ...PPX Series User Manual 70 Red The alarm of short circuit occurs from temperature measurement...
Страница 151: ...MENU CONFIGURATION 151 Communicati on Monitor Display The icon indicates Communication Monitor is activated...
Страница 172: ...PPX Series User Manual 172 Schematic PPX Trigger input signal Analog connector 12 10 B COM B COM 100k...