PPX Series User Manual
164
Connection- alt.
shielding
If the wire shield needs to be grounded at the
voltage source (EXT-V), then the shield cannot
also be grounded at the negative (-) terminal
output of the PPX power supply. This would
short the output.
PPX
EXT-V
Analog
connector
+
-
6
5
Output
Terminal
2 core shielded
wire or twisted
pair
Pin5 → EXT-V (+)
Pin6 → EXT-V (-)
Wire shield → EXT-V ground (GND)
Steps
1.
Connect the external voltage according to the
connection diagrams above.
2.
Set the CC Control as External V.
Also, set Output Type per application
and turn On Output Enable.
Page 108
3.
Press the Output key. The current
can now be controlled with the
External voltage.