WITE32 Release Notes Version 3.11

 

 

 

12.  In WDCP2002, when some spinstand tests are running, an attempt to load or unload Add-Ins from Add-Ins 

Manager dialog may cause several error messages. To avoid this the menu item 

File | Addin...

 that opens the 

Add-Ins Manager

 dialog box is disabled if one of the WDCP2002 tests is running. 

13. The 

Head Alignment

 test fails with unclear diagnostics if the service track for the 

Head Alignment

 test is written 

using a voltage read bias mode amplifier, and then you change the amplifier or the bias mode to the current 
mode, or vice versa.  

14.  The range of valid values for the 

Y Base Offset

 parameter in the 

Product Parameters 

dialog box of WDCP2002 is 

increased to –10 Inch ... +10 Inch range. 

15.  The time stamp of the spinstand errors shown in the WITE32 error messages is calculated wrongly. WITE32 

version 3.10 shows the time since the spinstand reset instead of the astronomical time. 

3.2 

Other Spinstands and Servo 

1.  For all Guzik 1701 and 312 spinstands, the 

Balancing

 procedure option N (balancing cap: 2.5) and option O 

(balancing cap: 1.8) shows instructions for 6 balancing screws instead of 3 balancing screws.  

2.  The Head Stack acoustic sensor emergency is handled wrongly on Guzik 1701B spinstands. When the acoustic 

sensor detects a scratch on the disk during Start Device, the spinstand immediately displays the 

Emergency 

Recovery Procedure

 dialog box and requires operator to unload a Head Stack manually. The new version of 

WITE32 stops operation, unloads a Head Stack normally, and then provides the proper diagnostic message 

"Acoustic Sensor detected"

.  

3. The 

PES Acquisition from Data

 method of the PES Analysis test is available only for spinstands equipped with 

Guzik Servo Revision 2 (i.e. Guzik S1701B and V2002 spinstands). In the new version of WITE32 this test is 
available for all spinstands.  

4.  The error message 

“No response from spinstand after IPL”

 occurs in the IPL Mode when Servo Controller or 

Motion Controller DSP code is corrupted in the spinstand flash memory. This bug affects 1701B spinstands only.  

5. The 

Servo Calibration

 dialog box shows the 

Track Width 

value instead of the 

Track Pitch 

value for Servo-1 

spinstands. 

6.  The erase step size validation in the 

Servo Erase

 operation test always fails. 

3.3 

Analog Front End 

1.  WITE32 does not report MR5 overheat error message 

“Head Amplifier Overheating Detected”

 when the MR5 

board is overheated. 

2.  The following bugs are fixed in the head amplifier drivers: 

• 

81G5114 – Bit 3 in the undershoot register (USC[3]) is set to 1 on start. 

• 

GTEMR5 – The amplifier P/N 23-321300-01 does not turn off power on Stop Device.

 

• 

GTEMR5 – the WDK function OvershootStepGet() always returns zero. 

• 

MR51796M – No readback from the chip registers. 

• 

MR5M – Any test (except for the Production test) started by pressing the 

Soft Button

 in the 

Operator 

Panel

 does not reload head amplifier zone parameters. As a result the MR5M head amplifier does not 

properly set the read bias and write current on start and fails the MR-Impedance measurement.

 

• 

MR5M – Intermittently fails to set read bias on start device (sets bias to unpredictable value). 

• 

PA7515, SR1774, SR1796 – WDK functions OvershootStepGet () and OvershootWidthStepGet() 
always return zero.

 

 

 

 

 

Guzik Technical Enterprises

 

25

 

Содержание WITE32

Страница 1: ...Release Notes WITE32 Version 3 11 6 28 2004 Copyright 2004 Guzik Technical Enterprises All rights reserved...

Страница 2: ...Configuration 8 2 1 4 Tests With Band Erase Configuration 10 2 1 5 Tests With Erase Preconditioning 12 2 2 New Adjacent Track Interference Module WATI 13 2 3 Spinstands and Servo 13 2 3 1 New Servo C...

Страница 3: ...version 3 11 an RWA 2000 series equipped with Servo Revision 3 is compatible with Guzik V2002 and Canon spinstands only The servo will not be available if you connect Servo 3 RWA to Servo 2 or Servo 1...

Страница 4: ...tor item is marked with the green check box it means that all the required hardware and cables are installed and compatible with Micro Actuator The list in the round brackets shows whether CW CCW or b...

Страница 5: ...5 SR1671 SR1850B SR1852 SR1880 TLS26A954AA TLS26D714 WABASH3x2 1 6 Head Stacks The following head stacks are initially supported in WITE32 version 3 11 81G214M MAGELLAN_81G214M PA7540B NOVA_PA7540B SR...

Страница 6: ...nt The new system options make the erase operation configurable 2 You can configure erase parameters in one dialog box and use this one common set of parameters in all tests and operations rather than...

Страница 7: ...values in mA used in the erase operation You can configure individual write current for the first and the second revolution The Use System check boxes If the check box is checked the Write Current te...

Страница 8: ...m erase or band erase operation setup or to override the system erase or band erase operation setup with the custom settings 2 1 2 Tests Without Erase Configuration The following tests perform erase o...

Страница 9: ...ample Original erase operation parameters Figure 3 SNR Test Configuration in WITE32 version 3 10 The following is the new SNR test configuration dialog box in WITE32 version 3 11 A new button opens a...

Страница 10: ...in earlier WITE32 versions Some test configurations have been modified to use System Band Erase Their original band erase operation configurations are converted to the new format Error Tests module P...

Страница 11: ...peration parameters Original erase operation parameters Figure 6 Low Erase Current Test Configuration in WITE32 version 3 10 The following is the new configuration dialog box in WITE32 version 3 11 A...

Страница 12: ...em Erase option is added to preconditioning selection of WITE32 tests When this option is selected the System Erase configuration will be used for preconditioning The following is the list of the test...

Страница 13: ...the head is stable is displayed once and the servo control is turned off If this check box is disabled WITE32 will keep the servo control enabled regardless of the quality of the signal from the head...

Страница 14: ...s direction of rotation and speed RPM device options disk working area skew angle parameters start and stop options The Heads tab page The page contains a number of parameters related to head geometry...

Страница 15: ...Inch 70 mm is now supported This disk type is added to the Product Parameters dialog box The Disk type 1 3 Inch is no longer supported The chuck design for 1 3 Inch disk never existed for V2002 spins...

Страница 16: ...gle and store the resulting value in the spinstand EEPROM The cartridge angle 0 or 10 degrees was added to the head loader angle in respect to the XY stage of the spinstand the V2002 head loader has a...

Страница 17: ...les or disables the hardware control of the Micro Actuator for the current product configuration If you choose to disable the device the hardware will not apply voltage to the Micro Actuator when the...

Страница 18: ...e cartridge for which the XY alignment is being performed Please refer to the section 2 3 2 3 for more details on the cartridge angle Figure 14 XY Alignment Test Configuration Dialog Box All Head Para...

Страница 19: ...t connect the Scope Point 7 output to the Index In input of RWA using a BNC to BNC cable and check the Enable Write With Retries Operation check box in the Measurement Options dialog box click the Con...

Страница 20: ...alibration regardless of the current cutoff frequency and the calibration signal spectral content This led to unpredictable measurement results in the case of changes of the cutoff frequency or the si...

Страница 21: ...e box see Figure 16 above In previous versions of WITE32 only the basic error information is shown in the More Info mode when the Errors at The Test End option is enabled 2 8 Head Selection in the WIT...

Страница 22: ...efer to the Micro Actuator Tests User s Guide P N 02 107264 02 for the details on the Micro Actuator operations 5 If an external module registration fails an error message is displayed to explain the...

Страница 23: ...0 and earlier generic internal names are used in the dialog box For example WrtFlux is used for the flux reversal value in the WriteFlux operation test of WITE32 version 3 10 which is renamed to Mflux...

Страница 24: ...if the destination position is reachable before the Y stage starts moving such that the Y stage does not hit the screw and the warning message is shown when the specified destination position is out...

Страница 25: ...ly and then provides the proper diagnostic message Acoustic Sensor detected 3 The PES Acquisition from Data method of the PES Analysis test is available only for spinstands equipped with Guzik Servo R...

Страница 26: ...metric and noisy 2 The band erase operation in the Overwrite test algorithm erases T 4 0 and T 4 offsets rather than T 2 0 and T 2 where T is the track size 3 The COV Detection Type in Popcorn measure...

Страница 27: ...the special 9999 values indicating the error when the measurements fail 8 Allocated sectored result memory in the read channel board for RWA 2000 series is smaller than on the RWA 2585 and RWA 2550 f...

Страница 28: ...WITE32 The table below lists the operating systems approved to work with WITE32 version 3 11 Operating system Supported in WITE32 version 3 11 Windows XP Service Pack 1a with Patch 822603 YES Windows...

Страница 29: ...TER 5 KNOWN ISSUES For Guzik V2002 spinstand after upgrading from WITE32 version 3 10 to WITE32 version 3 11 it is necessary to rewrite service track for the Head Alignment test otherwise execution of...

Отзывы: