WITE32 Release Notes Version 3.11

 

 

 

2.2 

New Adjacent Track Interference Module (WATI)  

The new Adjacent Track Interference module (WATI module) is added to WITE32 package. This module contains 
one test – the Adjacent Track Interference test.  Please refer to the 

WATI Module Users Guide

 (P/N 02-107268-02) 

for the description of the WATI module and the test. The WATI module requires a license and possible hardware 
upgrade; please contact 

[email protected]

 and provide them your RWA and Spinstand EEPROM Dump information 

to determine whether your test system requires an upgrade, and to request a quote for the license. 

2.3 

Spinstands and Servo  

2.3.1  New Servo Control Dialog Box  

The new 

Servo Control

 dialog box is added in WITE32 rev 3.11.  It can be opened by selecting the 

Control | Servo…

 

menu item in the WITE32 Engineering Mode Dashboard.  

Figure 9 

New Servo Control Dialog Box 

This dialog box has two check box controls: 

1. 

Disable Servo Control if Loop is Unstable.

 If this check box is enabled, WITE32 turns off the servo control 

in case the signal from the head becomes unstable. The warning message 

“Servo control is turned off. Check 

that the signal from the head is stable”

 is displayed once, and the servo control is turned off. If this check 

box is disabled, WITE32 will keep the servo control enabled regardless of the quality of the signal from the 
head. This may result in the recurring error messages during head positioning, however, the system will try 
to control the head position using servo whenever it is possible.  It is recommended to enable this option in 
the production environment, such that testing the heads producing poor signal can be finished in non-servo 
mode. This control is available only for Guzik V2002 spinstands. For all other spinstands this check box is 
disabled. 

2. 

MA Enable

 – This control is reserved for future use with Micro Actuator. 

2.3.2  V2002 Product Parameters Dialog Box Modifications  

The 

Product Parameters

 dialog box contains the parameters of the Head and the Media under test. For V2002 

spinstand, you can open the 

Product Parameters

 dialog box by selecting the 

Configure | Device…

 menu item in the 

WITE32 Engineering Mode Dashboard, pressing the 

Run Alignment Program

 button in the 

Device Configure

 dialog 

box, and pressing the 

Edit Product Parameters…

 button on the 

Equipment

 tab of the WDCP2002 program. 

The layout of the 

Product Parameters

 dialog box for V2002 spinstand is changed and new parameters are added in 

WITE32 version 3.11. 

 

 

 

 

Guzik Technical Enterprises

 

13

 

Содержание WITE32

Страница 1: ...Release Notes WITE32 Version 3 11 6 28 2004 Copyright 2004 Guzik Technical Enterprises All rights reserved...

Страница 2: ...Configuration 8 2 1 4 Tests With Band Erase Configuration 10 2 1 5 Tests With Erase Preconditioning 12 2 2 New Adjacent Track Interference Module WATI 13 2 3 Spinstands and Servo 13 2 3 1 New Servo C...

Страница 3: ...version 3 11 an RWA 2000 series equipped with Servo Revision 3 is compatible with Guzik V2002 and Canon spinstands only The servo will not be available if you connect Servo 3 RWA to Servo 2 or Servo 1...

Страница 4: ...tor item is marked with the green check box it means that all the required hardware and cables are installed and compatible with Micro Actuator The list in the round brackets shows whether CW CCW or b...

Страница 5: ...5 SR1671 SR1850B SR1852 SR1880 TLS26A954AA TLS26D714 WABASH3x2 1 6 Head Stacks The following head stacks are initially supported in WITE32 version 3 11 81G214M MAGELLAN_81G214M PA7540B NOVA_PA7540B SR...

Страница 6: ...nt The new system options make the erase operation configurable 2 You can configure erase parameters in one dialog box and use this one common set of parameters in all tests and operations rather than...

Страница 7: ...values in mA used in the erase operation You can configure individual write current for the first and the second revolution The Use System check boxes If the check box is checked the Write Current te...

Страница 8: ...m erase or band erase operation setup or to override the system erase or band erase operation setup with the custom settings 2 1 2 Tests Without Erase Configuration The following tests perform erase o...

Страница 9: ...ample Original erase operation parameters Figure 3 SNR Test Configuration in WITE32 version 3 10 The following is the new SNR test configuration dialog box in WITE32 version 3 11 A new button opens a...

Страница 10: ...in earlier WITE32 versions Some test configurations have been modified to use System Band Erase Their original band erase operation configurations are converted to the new format Error Tests module P...

Страница 11: ...peration parameters Original erase operation parameters Figure 6 Low Erase Current Test Configuration in WITE32 version 3 10 The following is the new configuration dialog box in WITE32 version 3 11 A...

Страница 12: ...em Erase option is added to preconditioning selection of WITE32 tests When this option is selected the System Erase configuration will be used for preconditioning The following is the list of the test...

Страница 13: ...the head is stable is displayed once and the servo control is turned off If this check box is disabled WITE32 will keep the servo control enabled regardless of the quality of the signal from the head...

Страница 14: ...s direction of rotation and speed RPM device options disk working area skew angle parameters start and stop options The Heads tab page The page contains a number of parameters related to head geometry...

Страница 15: ...Inch 70 mm is now supported This disk type is added to the Product Parameters dialog box The Disk type 1 3 Inch is no longer supported The chuck design for 1 3 Inch disk never existed for V2002 spins...

Страница 16: ...gle and store the resulting value in the spinstand EEPROM The cartridge angle 0 or 10 degrees was added to the head loader angle in respect to the XY stage of the spinstand the V2002 head loader has a...

Страница 17: ...les or disables the hardware control of the Micro Actuator for the current product configuration If you choose to disable the device the hardware will not apply voltage to the Micro Actuator when the...

Страница 18: ...e cartridge for which the XY alignment is being performed Please refer to the section 2 3 2 3 for more details on the cartridge angle Figure 14 XY Alignment Test Configuration Dialog Box All Head Para...

Страница 19: ...t connect the Scope Point 7 output to the Index In input of RWA using a BNC to BNC cable and check the Enable Write With Retries Operation check box in the Measurement Options dialog box click the Con...

Страница 20: ...alibration regardless of the current cutoff frequency and the calibration signal spectral content This led to unpredictable measurement results in the case of changes of the cutoff frequency or the si...

Страница 21: ...e box see Figure 16 above In previous versions of WITE32 only the basic error information is shown in the More Info mode when the Errors at The Test End option is enabled 2 8 Head Selection in the WIT...

Страница 22: ...efer to the Micro Actuator Tests User s Guide P N 02 107264 02 for the details on the Micro Actuator operations 5 If an external module registration fails an error message is displayed to explain the...

Страница 23: ...0 and earlier generic internal names are used in the dialog box For example WrtFlux is used for the flux reversal value in the WriteFlux operation test of WITE32 version 3 10 which is renamed to Mflux...

Страница 24: ...if the destination position is reachable before the Y stage starts moving such that the Y stage does not hit the screw and the warning message is shown when the specified destination position is out...

Страница 25: ...ly and then provides the proper diagnostic message Acoustic Sensor detected 3 The PES Acquisition from Data method of the PES Analysis test is available only for spinstands equipped with Guzik Servo R...

Страница 26: ...metric and noisy 2 The band erase operation in the Overwrite test algorithm erases T 4 0 and T 4 offsets rather than T 2 0 and T 2 where T is the track size 3 The COV Detection Type in Popcorn measure...

Страница 27: ...the special 9999 values indicating the error when the measurements fail 8 Allocated sectored result memory in the read channel board for RWA 2000 series is smaller than on the RWA 2585 and RWA 2550 f...

Страница 28: ...WITE32 The table below lists the operating systems approved to work with WITE32 version 3 11 Operating system Supported in WITE32 version 3 11 Windows XP Service Pack 1a with Patch 822603 YES Windows...

Страница 29: ...TER 5 KNOWN ISSUES For Guzik V2002 spinstand after upgrading from WITE32 version 3 10 to WITE32 version 3 11 it is necessary to rewrite service track for the Head Alignment test otherwise execution of...

Отзывы: