17
Steps
Motor
Contra-angle handpiece
Drying
Drying should be conducted after cleaning. Wipe
off the residual water on the motor surface with
dry absorbent soft cloth.
Drying of outside of instrument
through drying cycle of
washer-disinfector. If necessary,
additional manual drying can be
performed through lint free towel.
Insufflate cavities of instruments
by using sterile compressed air.
If your washer-disinfector does not
have an automatic drying function,
please dry the device after cleaning
and disinfection. The drying
method as below:
1) Spread a clean white paper
(white cloth) on the flat table, place
the contra-angle handpiece on the
white paper (white cloth), and then
dry the contra-angle with filtered
dry compressed air (maximum
pressure 3 bar). When no liquid is
sprayed on the white paper (white
cloth), it indicates that the contra-
angle handpiece is completely dry.
2) The device can also be dried
directly in a medical drying cabinet
(or oven).The recommended
drying temperature is 80℃ and
the time should be 15 minutes.
Note:
1).Dry the contra-angle handpiece
repeatedly if necessary (refer to
section "Drying").
2).The air used for drying must be
filtered by HEPA.
3) The device should be dried in a
clean area.
5)The drying temperature
should not exceed 137 °C