6 Safety precautions
Cautions
:
6.1 For repairs and purchase of spare parts, please contact our
authorized supplier.
6.2 The accuracy of the speed monitoring depends on the high
precision performance of the handpiece installed on the micro
motor. If the handpiece of other manufacturers is used, the actual
speed value may not be displayed correctly. To ensure the correct
display of the actual speed, please use the matching handpiece.
6.3 Read this operating manual before use and fully understand the
functions of each part.
6.4 Check the operating status of the dental electric motor before
use to confirm that there is no abnormality.
6.5 Test the dental electric motor before use to ensure accurate
operation.
6.6 If the motor is malfunctioning permanently (excessive vibration,
much noise and heat generation, etc.), please switch off immediately
and return it to the authorized dealer.
6.7 Clean the control panel with a damp cloth and turn off the
power before cleaning.
7 Cleaning, disinfection and sterilization
7.1 Motor and Contra-angle handpiece
Steps
Motor
Contra-angle handpiece
Warnings:
1. Do not spray any cleaning solution or lubricant
to the motor.
2. Do not use an automatic washer-disinfector to
clean the motor.
3. Except for the motor, the other parts such as
main unit, power adapter and tail are not
allowed to be sterilized.
Warnings:
1. The contra-angle handpieces are
not suitable for ultrasonic cleaning.
2. The contra-angle handpiece
should not be immersed in strong
detergents or disinfectants (alkaline
pH>9 or acid pH <5).
Resistance
to sterilizing
procedure
The motor could be capable of being subjected to
250 cycles of sterilizing procedure.
The contra-angle handpiece could
be capable of being subjected to
250 cycles of sterilizing procedure.