15
Steps
Motor
Contra-angle handpiece
Preparation at
the Point of Use
Remove the motor from the motor tail, and wipe
the surface of the motor with a soft cloth soaked
in cold water(<40°C).
The post-operative process must
be carried out immediately, no
later than 30 minutes after the
completion of the operation. The
steps are as follows:
Disconnect the contra-angle
handpiece from Motor.
Remove gross soiling of the
instrument with cold water
(<40°C) immediately after use.
Don’t use hot water (>40°C)
as this can cause the fixation of
residuals which may influence the
result of the reprocessing
process.
Transportation
The motor and contra-angle handpiece should be safety stored and transported to the
point of reprocessing to avoid any damage
and environment pollution.
Preparation for
reprocessing
The products must be reprocessed in a disassembled state.
Disassemble the contra-angle handpiece, motor and motor tail separately according
to section 3.5.
Pre-cleaning
Tools: tray, syringe, hose, clean and dry soft
cloth.
①
Use a syringe to draw 500ml of purified water
to flush the waterway and atomization airway.
Do not clean the cooling air circuit,
otherwise it will
damage the motor.
②
Wipe the water stains on the interface with
clean cloth.
③
Rinse the motor under running water for 2
min.
④
Rinse the motor surface with 500ml purified
water.
Tools: tray, soft brush, clean and
dry soft cloth
.
Perform manual pre-cleaning
until the handpiece is visually
clean. Rinse the chuck of bur
with running water for at least 10
seconds. Clean the surface with a
soft bristle brush.
Note: The water temperature
should not exceed 40℃ during
the washing stage, otherwise
the protein will solidify and be
difficult to remove.