11
screwdriver, then remove the battery cover, next remove the old battery,
replace it with a new one, and finally cover the battery cover and tighten
the screws.
Warning:
Improper replacement of lithium batteries may result in unacceptable
risks,so replacement of lithium batteries requires trained personnel.
4.5 Installation and disassembly of pushing ram
When pulling out the pushing ram, the limit switch must be kept
pressed.
The pushing ram can only be plugged in or unplugged from tail part of
the Hot Melting and Filling Instrument.
5 Operation method
Note:
During use, please do not contact the heating part of the Hot
Melting and Filling Instrument. Before use, remember to install the
Thermal Protector Cap to prevent users or patients from scalding.
1. Choose Gutta Percha Injecting Needle
Choose suitable Gutta Percha Injecting Needle (20ga, 23ga or 25ga)
according to the situation of patient. And tighten the Gutta Percha
Injecting Needle and handpiece (Note: not too tight). When using, the
Gutta Percha Injecting Needle can rotate to suitable angle within the
range of 360°in clockwise direction and counterclockwise direction. And
you can also use wrench to pre-bend the needle and adjust it to a better
operation angle as per your needs.
Warning:
When install the Gutta Percha Injecting Needle, please be sure that
the device is off and the head part of the device is cooling down. (About
5 minutes after shutdown of the Hot Melting and Filling Instrument, the
head part of it can cool down to the temperature that allows people to
touch.)