1.7.2 Input air pressure: 3.5bar to 4.5bar (350-450kPa)

1.7.3 Working mode: continuous operation, Intermittent operation

1.7.4 Weight: About 0.150kg

1.7.5 Connector: Meets the requirements of YY / T 0514

1.7.6 Operation environment:

a)Environment temperature: + 10℃~ +40℃ 

b)Relative humidity: 30% ~ 75% 

c)Atmospheric pressure:50kPa ~ 106kPa

2. Installation and operation

2.1 Schematic diagram 

2.1.1 Schematic diagram of Dental Air Polisher

Figure 2 Product schematic diagram

2.1.2 Schematic diagram of handpiece connection

Содержание AP-H

Страница 1: ...Dental Air Polishing Handpiece ...

Страница 2: ...Contents Preface 3 1 Product introduction 3 2 Installation and operation 7 3 Product function and operation 11 4 Cleaning disinfection and sterilization 20 5 Storage and transportation 26 6 Symbols 27 ...

Страница 3: ... Features of this device a Both Supragingival and subgingival air polishing are available enabling all round biofilm removal and periodontal treatment b Anti suction design prevents powder and moisture from being sucked back and prevents cross infection and dental unit interface contamination c Three section design enables ease of loading and unloading cleaning and maintenance d The tail cord conn...

Страница 4: ...4 1 2 Model and specifications 1 2 1 Model AP H AP H Plus 1 2 2 Specifications Length 177mm Width 52mm Height 88mm 1 3 Configuration Diagram of product external structure and accessories Figure 1 ...

Страница 5: ...5 Figure 1 Schematic diagram of structure ...

Страница 6: ...igment and for implant maintenance 1 6 Contraindications 1 6 1 The hemophilia patient is forbidden to use this equipment 1 6 2 The patients with heart pacemaker are forbidden to use this equipment 1 6 3 The doctors with heart pacemaker are forbidden to use this equipment 1 6 4 Heart disease patients pregnant women and children should be cautious to use the equipment 1 6 5 Patients with respiratory...

Страница 7: ...5 Connector Meets the requirements of YY T 0514 1 7 6 Operation environment a Environment temperature 10 40 b Relative humidity 30 75 c Atmospheric pressure 50kPa 106kPa 2 Installation and operation 2 1 Schematic diagram 2 1 1 Schematic diagram of Dental Air Polisher Figure 2 Product schematic diagram 2 1 2 Schematic diagram of handpiece connection ...

Страница 8: ...220kPa Temperature Up to40 Air supply pressure Adjust the air supply pressure of the external dental equipment to obtain the air supply pressure of 3 5 bar to 4 5 bar 350 450kPa Note Excessive air pressure will cause cracks or ruptures on the powder storage chamber and powder storage chamber cover of the Dental Air Polisher and even cause harm to the human body ...

Страница 9: ...rface and the handpiece pipeline interface with an air gun directly insert the Dental Air Polisher interface into the handpiece pipeline and then tighten the nut of the pipeline Note Fully dry the Dental Air Polisher interface and the handpiece pipeline interface before connection so as to avoid the spray head blockage from damp powder during use ...

Страница 10: ...piece connection After drying the inside of the handpiece and the Dental Air Polisher interface directly insert the handpiece into the handpiece interface of the gun for connection Note Fully dry the Dental Air Polisher interface and the handpiece ...

Страница 11: ...during use Figure 5 Handpiece connection 3 Product function and operation 3 1 Powder loading After connecting the Dental Air Polisher unscrew the powder storage chamber cover of the gun put the powder into the chamber clear the powder at the thread position of the powder storage chamber cover and ...

Страница 12: ...12 then tighten the cover Note Only woodpecker brand powder can be used and the amount of the powder added should not exceed the maximum scale mark Figure 6 Powder loading 3 2 Product operation ...

Страница 13: ...2 1 2 Generally the handpiece is held in a pen holding position 3 2 1 3 When using an Dental Air Polisher for normal air polishing it is recommended that the spray head air outlet and the tooth surface be kept at a distance of 3 5mm and the air polishing direction is recommended to be at an angle of 30 60 with the tooth surface as shown in Figure 7 Figure 7 Schematic diagram of subgingival air pol...

Страница 14: ...olding position 3 2 2 4 When using the subgingival handpiece for normal subgingival air polishing it is recommended to use a nozzle to remove the bioflm in the periodontal pocket at subgingival 4 9mm and perform up and down motion for scaling 3 2 2 5 The air polishing time of each periodontal pocket is recommended for no more than 5 seconds 3 2 2 6 Please ensure that the handpiece has a normal wat...

Страница 15: ...powder can be used for subgingival air polishing 3 2 3 Instructions for main components 3 2 3 1 Handpiece Working part the spray head of which can be pulled out by rotation In case of blockage the user can rotate the spray head to loosen and pull out it and then use the needle to negotiate The spray head can be sterilized at high temperature ...

Страница 16: ...enance 3 3 1 Powder clearing Open the powder storage chamber cover and clear the remaining powder on it Blow off the powder remaining in powder storage chamber with air Wipe off the powder residues on the surface of the Dental Air Polisher with alcohol cotton or dry towel Figure 9 Schematic diagram of powder clearing ...

Страница 17: ...out it as shown in Figure 10 Aim the air gun at the nozzle and blow out the powder inside the nozzle Use an air gun to blow the front and rear ends of the handpiece to clean the remaining moisture on the handpiece If the nozzle is blocked please use the needle to clear it Figure 10 Schematic diagram of handpiece cleaning ...

Страница 18: ... otherwise it will affect the product sealing and normal operation Keep the Dental Air Polisher dry Rain moisture and various liquids may damp the powder and affect its normal use Do not throw beat or vibrate the Dental Air Polisher Rough treatment of the gun will damage the strength and sealing of equipment Do not paint the instrument with pigment which will leave debris in the detachable parts a...

Страница 19: ... warm water and dry it Clogged powder storage chamber Check whether the holes on the air inlet pipe and powder outlet pipe in the powder storage chamber are blocked If blocked negotiate with a needle and blow off with an air gun No air coming out of the handpiece The waterway was not correctly installed Check whether the air source switch or airway of the dental unit is properly connected Clogged ...

Страница 20: ...proof O ring Replace waterproof O ring 4 Cleaning disinfection and sterilization Warnings The use of ultrasonic cleaning equipment and strong cleaning disinfectant alkaline pH 9 or acidic pH 5 will shorten the service life of the product In this case the manufacturer is not liable Do not expose the equipment to the high temperature above 138 Processing restrictions Only the handpiece and spray hea...

Страница 21: ...ter effective cleaning and disinfection can effective sterilization be carried out Please ensure that as part of the sterility of the product in use only fully validated equipment and product specific procedures are used for cleaning disinfection and sterilization and that validated parameters are followed during each cycle Please also comply with the applicable legal requirements of your country ...

Страница 22: ...hen rinse the dirt on the air polishing handpiece and spray head surface with pure water or distilled water deionized water 4 1 2 3 Dry the surface of the handpiece with a clean soft cloth and place it in a clean tray Note The water used here must be pure water distilled water or deionized water 4 2 Cleaning Cleaning automatic 4 2 1 The detergent shall be CE certified to be effective according to ...

Страница 23: ...sidues should meet cytotoxicity test requirements Note Please write down the selected cleaning equipment the steps and parameters of the processing program and the detergent 4 3 Disinfection 4 3 1 Place the handpiece and spray head of the Dental Air Polisher carefully in the disinfection box Contact between the handpieces is prohibited 4 3 2 Use a suitable rinsing adapter and connect the internal ...

Страница 24: ... temperature shall not exceed 20 minutes c The disinfection cycle shall comply with EN ISO 15883 4 4 Drying If the automatic drying function is not included in your cleaning and disinfection process please perform drying after cleaning and disinfection 4 4 1 Lay a clean white cloth on the workbench align the handpiece and spray head with the cloth and then blow the handpiece and spray head with fi...

Страница 25: ...cted and dried handpiece and its accessories and quickly package them in a medical sterilization package or an exclusive holder a sterile box Note 1 The packaging used shall conform to ISO 11607 2 It can withstand the high temperature of 138 and has sufficient steam permeability 3 The packaging environment and related tools must be cleaned regularly to ensure cleanliness and prevent the entry of p...

Страница 26: ...lization time at 134 C is 20 minutes 5 Storage and transportation 5 1 Storage 5 1 1 The product should be handled with care and away from the earthquake source Ensure that it is installed or stored in a cool dry and ventilated place indoors 5 1 2 Do not mix the product with toxic corrosive flammable or explosive materials during storage 5 1 3 The product should be stored in the environment with re...

Страница 27: ... be clean and be disinfected regularly 2 Product storage must be batched marked and recorded 5 2 Transportation 5 2 1 Avoid excessive impact and vibration during transportation Handle it with care and avoid inversion 5 2 2 It should not be mixed with dangerous goods during transportation 5 2 3 Avoid sunlight rain or snow during transportation 6 Symbols Date of manufacture Follow Instructions for U...

Страница 28: ...28 Handle with care Keep dry Appliance complies with WEEE directive Can be autoclaved Temperature limit for storage Atmospheric pressure for storage Maximum sand volume Humidity limit for storage ...

Страница 29: ......