EN ISO 15883. For specific steps, see the automatic disinfection part in the
next section, “Disinfection”.
[Note]:
a)The detergent used here must be compatible with the Dental Air
Polisher and only freshly prepared solutions can be used.
b)At the washing stage, the water temperature should not exceed
45° C; otherwise the protein will solidify and be difficult to remove.
c)After cleaning, chemical residues should meet cytotoxicity test
requirements. Note: Please write down the selected cleaning equipment,
the steps and parameters of the processing program, and the detergent.
4.3 Disinfection
4.3.1 Place the handpiece and spray-head of the Dental Air Polisher
carefully in the disinfection box. Contact between the handpieces is
prohibited.
4.3.2 Use a suitable rinsing adapter and connect the internal water pipe
to the washer-disinfector.
4.3.3 Start the program
4.3.4 After the procedure is completed, remove the handpiece and
spray-head from the washer-disinfector for inspection (see the “Daily