drying time should be 15 〜 40 minutes.
4.4.3 After drying, remove the handpiece and spray-head from the
drying oven for testing (see “Daily maintenance”).
[Note]:
a)Product drying must be done in a clean place.
b)Drying temperature shall not exceed 138℃ ;
c)The equipment in use should be inspected and maintained regularly.
4.5 Packaging
Assemble the disinfected and dried handpiece and its accessories, and
quickly package them in a medical sterilization package (or an exclusive
holder, a sterile box).
[Note]:
(1)The packaging used shall conform to ISO 11607;
(2)It can withstand the high temperature of 138℃ , and has sufficient
steam permeability;
(3)The packaging environment and related tools must be cleaned
regularly to ensure cleanliness and prevent the entry of pollutants;
(4)Avoid touching different metal parts when packing.
4.6 Sterilization
Only the following steam sterilization procedures (Fractionated pre-