Section 2
OM6624CT-3000-5-00
11 May, 2016
2-6
Step 2)
Turn on the components of the system in the following order:
-
The 6623A-3S Control Unit.
-
The DVM.
-
The 66259 Extender Control Unit if implemented.
-
The CS1000 Current Sources, if implemented, up to three units only; or the
separate current source.
Step 3)
Zero the DVM reading then set a nominal positive test current of 100 A.
Step 4)
Observe and record the voltage on the potential terminals of the reference shunt as
displayed on the DVM after the test current has stabilized for a few minutes.
Observe that the balance meter on the front panel returns to and remains near the
centre line within the graduation lines of the meter. A small offset is a satisfactory
indication that a balance condition exits on the 6624CT-3000 Transformer Unit.
Note that an averaging function on the DVM will be required to smooth out small
variations in the current source output.
Step 5)
Set a negative nominal test current of 100 A.
Step 6)
Observe and record the voltage on the potential terminals as displayed on the DVM
after the test current has stabilized for a few minutes
.
Step 7)
Divide the absolute value of the mean of the two voltages, and divide the result by
the known calibrated resistance value of the reference shunt.
Step 8)
Multiply the result by the range factor of 1000 and check that the result is within
0.1 % of the 100 A test current requested.
Note: The test should execute with no measurement errors indicated. This is not the 6624CT-
3000 full rated accuracy but serves as a functional test for the purposes of determining the
instrument has been received in proper working order.
Note: It is recommended to keep the CS1000 modules powered off when the instrument is not
in use for extended periods of time. Always turn the CS1000 modules off first then the
remaining components.