2.4.
6624CT-3000 REAR VIEW
Figure 2-3 : 6624CT-3000 Rear View
2.4.1.
Line Input Connectors
The
Line Input Connectors
are a combination of a 3-prong AC standard male IEC
connector, and an ON/OFF switch. The 6623A-3S Controller has a fused line input with
access for fuse replacement. The CS1000’s have a 15 A circuit breaker as part of the line
input connector. The line input connectors accommodate nominal voltages of 208 to 240 V.
2.4.2.
Control Connector
The
Control Connector
of the 6623A-3S is a 7 pin circular connector which is used to
connect the Model 66259 Control Unit. The 66259 provides program signals for the
CS1000 Current Sources and the 6623A-3S tracking of the current comparator.
2.4.3.
DCCT-3000 Connector
The
DCCT-3000 Connector
is a 16 pin circular connector which is used to connect the
necessary signals to the Model 6624CT-3000 Current Transformer 'CONTROL'
connector.
Only 12 of the pin connections are utilized.
2.4.4.
CS1000 Connector
The
CS1000 Connector
is a 7 pin circular connector which is used to connect the necessary
signals to the first 1000A current source 'INPUT' connector. The 'OUPUT' connector of first
CS1000 is connected to the 'INPUT' of the next unit and then to the third unit.
Note that the 'OUTPUT' of the last or upper CS1000 module must have a termination
plug, (GPN 31061-02-09) installed for proper operation.