Section 5
OM6624CT-3000-5-00
11 May, 2016
5-1
5. TROUBLESHOOTING AND MAINTENANCE
5.1.
INTRODUCTION
This section covers troubleshooting, and maintenance of the 6624CT-3000 High Current
DC Transformer.
5.1.1.
Model 6623A-3S Controller Fault Indications
The meter indicator on the 6623A-3S Controller front panel will remain off scale whenever
a large imbalance condition is detected in the comparator circuit. This may be due to
improper connections to the range extender, improper measurement setup or an actual
circuit failure. Under extreme unbalance conditions the RED 'Power' indicator may dim. In
this case it is important to turn off the test current and turn off the 6623A-3S to avoid
damage to the electronic circuits.
Check that a shorting link is installed on the unused Burden terminals and a termination
plug is installed on the 'Control' connector on the 6623A-3S if the Model 66259 Extender
Control Unit is not used.
5.1.2.
Maintenance
Preventive maintenance is limited to checking the instrument operation, test current ratio
accuracy, cooling fan operation and making sure that the air filter screens are clean. The air
filter screens should be removed and cleaned on a yearly basis.