Troubleshooting
The chapter describes the common problems with thermometer Web Sensor
AR003826
, Web Sensor
AR005075
, Web Sensor
AR003827
and methods how to fix these
problems. Please read this chapter before you will call technical support.
I forgot the device IP address
IP address is factory set to 192.168.1.213. If you had changed it and forgot new IP address,
run the TSensor software and press Find device... In the window are displayed all
available devices.
I cannot connect to the device
In search window is only IP and MAC address displayed
Other details are marked N/A. This problem occurs if IP address of the device is set to
another network.
Select the window Find device in TSensor software and press Change IP address.
Follow the software instructions. To assign IP address automatically using DHCP server,
set the device IP address to 0.0.0.0.
Device IP address is not displayed in window Find device
In TSensor software menu press Help! My device was not found! in window Find
device. Follow the software instructions. MAC address of the device can be found on
product label.
The device is not found even after manually setting
MAC address
This problem occurs especially in cases when the IP address of the device belongs to another
network and also Subnet mask or Gateway are incorrect.
In this case is DHCP server in the network necessary. In TSensor software menu press
Help! My device was not found! in window Find device. As new IP address set 0.0.0.0.
Follow the software instructions. An alternative is to reset device to factory defaults using
factory-defaults button.