PROPERLy STORING COBOSAFE-CBSF
40
CoboSafe-CBSF
325-2810-012US15
13.5.3
Temperature/Humidity
The integrated thermometer and hygrometer are calibrated along with the CBSF measuring
devices.
If calibration is not possible for various reasons, an external suitable measuring instrument must
be used to determine the parameters.
13.5.4 K1 Damping elements
The damping elements K1 are checked by the manufacturer for compliance with their specifi-
cation during calibration of the CBSF-XS.
Otherwise, the damping elements K1 should be replaced when worn or at the latest once a
year.
13.5.5 Spare Parts
Use only original spare parts or parts approved by the manufacturer. Spare parts can be
obtained from GTE Industrieelektronik GmbH. Please contact customer service for assistance.
13.5.6 Firmware Update
Firmware is updated via the CoboSafe-Vision software (this does not apply to CBSF-Basic).
14 Properly Storing CoboSafe-CBSF
■
Ensure prescribed storage conditions (Refer to chapter “Technical Data”)
■
Always store CBSF-XS devices in the included transport cases.
■
Store CBSF-XS devices such that they are not exposed to shaking or vibrations.