7.4 Identification

The following device variants of the DID are available. Regarding 
detailed information of the device variants please refer to the 
technical data located at the end of the manual:

7.4.1 Nameplate

The control unit CU 382 is typified by a nameplate, as shown on 
the right, that contains the following information:

Fig. 5

Nameplate

7.4.2 Type key

Example: DID-3 BF3-FCL2/TCL2/pH

T

M

06

 59

20

 03

16

Pos. Description

1

Manufacturer‘s name and country of origin

2

Item number (type)

3

Max. numbers of sensor

4

Bar code

5

Device serial number (S/N)

6

Enclosure class

7

Several certification marks

8

QR code

9

Voltage [V]

10

Frequency [Hz]

11

Power consumption [W]

12

Acceptable humidity limits

13

Acceptable temperature limits

1

2
3

5

7

8

9, 10, 11

12

13

4

6

Code

Description

Control unit CU 
382 variants

DID-1

CU 382-1 for 1par 
temperature, 100-240 VAC, 50-60 Hz

DID-3

CU 382-3 for 3 para 
temperature, 100-240 VAC, 50-60 Hz

Installation 
variant

BF1

Bypass flow cell, 1 sensor

BF3

Bypass flow cell, 3 sensors

TI1

Tank immersion clean water, 1 sensor

TI2

Tank immersion clean water, 2 
sensors

Parameters/
Sensors

FCL2

Free chlorine 0-2 ppm

FCL20

Free chlorine 0-20 ppm

TCL2

Total chlorine 0-2 ppm

TCL20

Total chlorine 0-20 ppm

CDI2

Chlorine dioxide 0-2 ppm

CDI20

Chlorine dioxide 0-20 ppm

HP2

Hydrogen peroxide 0-200 ppm

HP20

Hydrogen peroxide 0-2000 ppm

PA2

Peracetic acid 0-200 ppm

PA20

Peracetic acid 0-2000 ppm

pH

pH 2-12 pH up to 10 bar and 70 °C

ORP

ORP, -2000 mV to +2000 mV up to 10 
bar and 70 °C

CND

Conductivity, 1-500000 µS/cm (2-42 
PSU), up to 20 bar and 70 °C 

Remarks:
• All sensors include temperature measurement
• Disinfectant parameters (FCL,TCL, CDI, HP or PA) are only possible 

with BF1 or BF3

• BF1 and BF3 include flow detection
• BF1 and BF3 systems include 1.0 m sensor cables
• Installation variants TI1 and TI2 are only possible for parameters PH, 

ORP or CND

• TI1 and TI2 systems include 7.5 m sensor cable(s)
• Mains cable is generally not included, please select and order from 

accessories list

Содержание DID

Страница 1: ...DID Installation and operating instructions GRUNDFOS INSTRUCTIONS...

Страница 2: ...vent of product failure 3 1 5 Working with chemicals 3 2 Safety instructions 3 3 Receiving the product 3 3 1 Transporting the product 3 3 2 Scope of delivery 3 3 3 Product updates other 3 4 Installing...

Страница 3: ...nsor installation fitting control unit etc No external interferences i e no electric and electro magnetic interferences by leakage current earth fault of pumps electric motors high voltage currents et...

Страница 4: ...ll them out by temporary increase of the discharge 5 Starting up the product 5 1 Initial startup Once the mounting and installation have been completed and checked the initial start up of the DID will...

Страница 5: ...the selection of the language see section 9 2 8 Select language Subsequently reading and information display will be shown The most recent information obtained by the CU 382 will be displayed even wh...

Страница 6: ...cludes a PID Proportional Integral Derivative controller which is capable of dosing disinfectants to the monitored water drinking bathing waste water The hydraulic part of the station manages the comp...

Страница 7: ...ers temperature 100 240 VAC 50 60 Hz Installation variant BF1 Bypass flow cell 1 sensor BF3 Bypass flow cell 3 sensors TI1 Tank immersion clean water 1 sensor TI2 Tank immersion clean water 2 sensors...

Страница 8: ...Free chlorine dosing control with validation via ORP Free Chlorine 0 2 ppm ORP 2000 2000 mV DID 3 BF3 FCL2 ORP 98915663 Free chlorine dosing control with wide pH variation and validation via ORP Free...

Страница 9: ...cknowledges alarms in the alarm screen Right button Moves right in views or values Down button Moves one line downward the selected line has a triangle as first character Decreases values OK 1 2 3 4 5...

Страница 10: ...tons and pressing OK button will open up the Quick calibration menu Before taking a sample to perform a calibration rinse the sampling point of the flow cell at least 3 times to remove all residuals f...

Страница 11: ...e first row shows date and time the alarm has occurred the first time In the rows two to four the alarm text is displayed With OK button every alarm can be acknowledged If the reason for the occurrenc...

Страница 12: ...position 9 2 3 Cleaning device In the Cleaning menu a connected cleaning device can be configured Cleaning is only needed if one of the connected sensors is submersed and a automatic cleaning device...

Страница 13: ...he Parameter settings menu a parameter can be calibrated monitored linked to an alarm linked to an output and removed Also an overview of all settings of a parameter can be displayed 9 3 1 Calibrate e...

Страница 14: ...and Value will show the current calibrated value Offset and Slope will show the calibration coefficients which are stored onto the sensor Performing a Span calibration This type of calibration is an...

Страница 15: ...digital outputs see section 9 3 6 Output settings 9 3 6 Output settings In the Output settings menu a parameter can be allocated to an analog or digital output If an analog output is chosen 4 mA defi...

Страница 16: ...ght button Pressing OK button will confirm the maximum of pulses min 9 3 7 Parameter info The Parameter info screen shows a summary of all important settings and information for one parameter 9 3 8 pH...

Страница 17: ...with Up and Down buttons and by pressing OK button the integral part can be changed with Up Down Left and Right button Pressing OK button will confirm the new value for Ti To disable the integral part...

Страница 18: ...OK button will confirm the type None as a type will deactivate the controller Hysteresis defines the hysteresis for the 2 P controller The switching points for the controller are Setpoint Hysteresis...

Страница 19: ...available Due to the fact that the controller output range is from 0 to 100 mapping isn t necessary The only parameter which can be changed is the maximum of pulses min for 100 output To change the m...

Страница 20: ...data will be copied to the USB stick Delete logs will delete all system llogfiles To delete all logfiles select Delete logs with Up and Down buttons and by pressing OK button the logfiles will be dele...

Страница 21: ...artup Routine functional check Suspicion of DID malfunction Modification of DID e g integration of additional sensor or device Change of measuring location Depending on the application water compositi...

Страница 22: ...and flow cell can be destroyed if too much force is used Flow detector shows no flow Inlet valve is closed Inlet strainer is blocked Open inlet valve Clean inlet strainer Water is not visible within t...

Страница 23: ...mm 4 Data Units TI1 TI2 General data Number of connectable S CAN sensors 1 2 Min liquid temperature C 0 1 Max liquid temperature C 45 Min ambient temperature C 0 1 Max ambient temperature C 45 Hydraul...

Страница 24: ...r external mechanical relay switch Digital input for external mechanical relay switch Galvanic isolation kV none none Maximum input frequency Hz 100 100 Low input voltage V 7 7 High input voltage V 8...

Страница 25: ...al display 70 x 40 viewing area trans reflective white background Data transfer USB host mode support for data transfer with USB stick Supports FAT FAT32 USB host mode support for data transfer with U...

Страница 26: ...hout packaging With packaging Product number kg kg DID 1 BF1 FCL2 4 26 8 095 98915656 DID 1 BF1 CDI2 4 26 8 095 98915657 DID 1 BF1 PA2 4 26 8 095 98915658 DID 1 BF1 HP2 4 26 8 095 98915659 DID 3 BF3 F...

Страница 27: ...6 6025 0416 TM06 6039 0216 Fig 12 Dimension of control unit CU 382 for wall mounting back mm Fig 13 Dimension of sensor with holder mm pipe with external diameter 50 mm not included 198 137 5 3 30 185...

Страница 28: ...Avoid direct sunlight for proper operation Before installation and operation DID the instruction manual must be read Avoid drying up the sensors The sensors must always be submerged in water Avoid suc...

Страница 29: ...Documents as listed Type of product Measurement Process Control Name of product DID Model number D 320 GF1 230 D 320 GF3 230 Low Voltage Directive 2014 35 EU Standard used EN 61010 1 2010 RoHS2 2011 6...

Страница 30: ...30...

Страница 31: ...Building 679 5 Yeoksam dong Kangnam ku 135 916 Seoul Korea Phone 82 2 5317 600 Telefax 82 2 5633 725 Latvia SIA GRUNDFOS Pumps Latvia Deglava biznesa centrs Augusta Deglava iel 60 LV 1035 R ga T lr 3...

Страница 32: ...5263 The name Grundfos the Grundfos logo and be think innovate are registered trademarks owned by Grundfos Holding A S or Grundfos A S Denmark All rights reserved worldwide Copyright Grundfos Holding...

Отзывы: