GLDM 64.1 Technical Manual rev. 1.30 February
2020
Page 22 of 50
8.5.2. RZ Reset Zero
[ RPDO1,01 ]
This command cancels the SZ command and the zero reading reverts to that set by the CZ command during
calibration.
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
RZ
OK
Zero point CZ active
The GGLDM
responds to the RZ command with either OK or ERR. If OK is returned then the “zero
action performed” bit in the Device Status (IS) response will be set to “0”.
8.5.3. ST Set Tare
[ RPDO1,08 ]
This command will activate the net weighing function by storing the current weight value as a tare value.
The weight signal must be “stable” within the limits set by NR (No Motion Range) and NT (No Motion Time)
commands for the “signal stable” bit to be active and set tare command to be accepted.
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
ST
OK
Tare performed / Net operation
If the weight signal is “stable”, the response to the IS command (Device Status) will show the “signal stable” bit
active and the ST command will be accepted (OK). If the “signal stable” bit is not active, the ST command will
be rejected and the GLDM will respond with ERR (error).
8.5.4. RT Reset Tare
[ RPDO1,04 ]
This command resets the tare and the weighing signal returns to gross mode.
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
RT
OK
Tare de-activated / Gross operation
The GLDM
responds to the RT command with either OK or ERR. If OK is returned then the “tare active” bit in the
Device Status (IS) response will be set to “0”.
8.5.5. SP Set Preset Tare
[ SDO 2100 sub 17 ]
This command sets a preset tare value.
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
SP
T+000000
Tare value 0 (factory default
SP1000
OK
Setup tare value 1000d