GLDM 64.1 Technical Manual rev. 1.30 February
2020
Page 15 of 50
8.2.3. CI
Set Minimum Output Value
[ SDO 2300 sub 08 ]
This command is used to set up the minimum output value. Permitted values are from
– 999999 to 0.
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
CI
I
–
000009
Request: CI =
–9 d
CE
E+000017 (example)
Request: TAC counter CE17
CE17
OK
Calibration sequence active
CI
-
10000
OK
Setup: CI = -10000 d
This value will determine the point at
which the output will change to “uuuuuuu”, signifying under-range.
Note:
In bipolar applications (e.g. force- or torque measurements) this parameter defines the max. output value
for input signals with negative sign.
Factory default
: CI =
–
999999
8.2.4. MR Set Multi-range / Multi-interval
[ SDO 2300 sub 0D ]
This command is only relevant, if CM 2 > 0 or CM 3 > 0. Is this the case, then this command defines, if the
application is multi-range or multi-interval. Permitted values are 0 (Multi-interval) or 1 (Multi-range).
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
MR
M+00000
Request: MR = 0 (Multi-interval)
CE
E+00017 (example)
Request: TAC counter CE17
CE17
OK
Calibration sequence active
MR1
OK
Setup: MR = 1 (Multi-range)
Note:
Single range applications ignore this parameter.
8.2.5. DS Set Display Step Size
[ SDO 2300 sub 0C ]
This command allows the output to step up or down by a unit other than 1.
Permitted values are 1, 2, 5, 10, 20, 50, 100, 200 and 500.
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
DS
S+00002
Request: Step size 2
CE
E+00017 (example)
Request: TAC counter CE17
CE17
OK
Calibration sequence active
DS50
OK
Setup: Step size 50
Legal for trade applications allow for up to 10000 intervals. The allowed step size has to be considered.
8.2.6. DP Set Decimal Point Position
[ SDO 2300 sub 0B ]
This command allows the decimal point to be positioned anywhere between leftmost and rightmost digits of the
6-digit output result. Permitted values are 0, 1, 2, 3, 4, 5 and 6. Position 0 means no decimal point.
Factory default:
DP = 3
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
DP
P+00003
Request: Position of decimal point 3
CE
E+00017 (example)
Request: TAC counter CE17
CE17
OK
Calibration sequence active
DP0
OK
Setup: no decimal point