
DG Spin
Resolution of Incidences
12-5
-
If after a complete centrifugation batch the Cover has not been opened
to remove the cards already centrifuged and the
button is pressed
again to start a new centrifugation batch, the system emits some short
sounds and the Open cover alarm flashes. To start a new
centrifugation, it is necessary to open the Cover.
-
If there is any error during centrifugation: Imbalances, centrifugation
speed out of range, incorrect function of out-of-balance control systems,
fault in the Cover opening system, excess current consumption of the
motor, etc. the
DG Spin
centrifuge emits some short sounds and the
corresponding indicator is displayed.
CAUTION!
To stop the Acoustic alarm, press
or
,
as appropriate. If the Acoustic alarm continues,
contact your authorised Technical Service.