DG Spin
Resolution of Incidences
12-4
as described in section 4.1.6 ”Security
ring”.
Restart again and if the error is detected
again, contact the Technical Service.
12.2
Error messages
Apart from the alerts and errors mentioned, the equipment may detect other error
conditions. In these cases, the equipment will carry out the following actions:
-
Stop the centrifugation in execution.
-
Ignore Buttons
,
and
.
-
Activate the Acoustic alarm with short sounds.
-
Activate the indicator
.
-
Show an
E
on the Display with the corresponding error code.
CAUTION!
If the
DG Spin
centrifuge shows an error
message in the Display, restart the equipment.
If the error persists, contact the Technical
Service and indicate the error code.
12.3
Acoustic alarm
The
DG Spin
centrifuge has an Acoustic alarm that can be programmed on three
levels and when configured it is activated with the volume selected by the
OPERATOR in the following situations:
-
Once the centrifugation batch in execution is completed. The acoustic
indicator will be activated with a long and continuous sound.
-
When a centrifugation batch is stopped by keeping
pressed down.
The
DG Spin
centrifuge emits three short sounds and a long sound
indicating that centrifugation has stopped at this time. The timer will
show by flashing, the value indicated at that time.
-
After a power cut during a centrifugation batch. When it works again,
the
DG Spin
centrifuge will emit short sounds and the timer will show,
by flashing, the value indicated at that moment.