![Greyline ISM 5.0 Скачать руководство пользователя страница 7](http://html.mh-extra.com/html/greyline/ism-5-0/ism-5-0_user-manual_2264645007.webp)
Page 5
Accuracy
± 1.0% of reading from 2 to 20 ft/sec
± 0.02 ft/sec below 2 ft/sec
Sensing Method
Electromagnetic, no moving parts
Pipe Size Range
3” through 72” nominal
Input Power
20 - 28 VAC 50/60Hz, 250 mA maximum
20 - 28 VDC, 250 mA maximum
Liquid Temperature Range
15° to 250° F
Ambient Temperature Range
-5° to 150° F
Operating Pressure
400 PSI maximum
Pressure Drop
Less than 0.1 psi at 12 ft/s velocity in 3” and larger pipes
Materials of Construction
Wetted metal components - 316L stainless steel
Sensor head - Polystyrene
Electronics enclosure - Powder coat painted cast aluminum
Enclosure Rating
Watertight, airtight, dust proof powder coated cast aluminum.
Electrical Connections
25’ of PVC jacketed cable with ½” NPT conduit connection
Output Signal(s)
Analog output (Isolated), selectable 4-20 mA, 0-10 V or 0-5 V
Frequency output, 0-15 V peak pulse, 0-500 Hz maximum
Scalable pulse output, isolated dry contact, contact rating 50 VDC @ 100 mA
maximum, pulse duration: 0.5, 1, 2 or 6 seconds
This product is covered by one or more of the following patents: 6,431,011 and 6,463,807.
All GREYLINE insertion type meters can be installed and removed via
a 1” or larger full port ball valve without system shutdown. The terms
“Standard” and “Hot Tap” refer to the installation method of the
isolation valve kit only.
Standard Installation Hardware: For new construction or scheduled
shutdown; once kit is installed, the flow meter can be installed
or removed without system shutdown.
Hot Tap Installation Hardware: For applications which require the
access hole in the pipe to be drilled through the valve using a
wet tap drilling machine while the hydronic system is
pressurized and operating.
NOTE: Installation hardware materials vary greatly based on pipe
material, pipe size and standard vs. hot tap versions.
1.4 ADDITIONAL REQUIRED HARDWARE
1.3 STANDARD FEATURES AND SPECIFICATIONS