Page 12
PDFM 5.1 Portable Doppler Flow Meter
SPECIAL FUNCTIONS
Language
Select
English
,
French
or
Spanish
Backlight
Select
High
,
Medium
or
Low
for continuous
backlight.
Select
Key Hi/Lo
for high backlight (for 1 minute)
after a keypress and then
Lo
backlight until a key is
pressed again.
Select
Key High
,
Med
or
Low
for backlight after
a keypress and then backlight off until a key is pressed
again.
Reset Totalizer
Press
and select
Yes
to erase and restart the
totalizer at zero.
Negative Totals
Select
Yes
to have reverse flow readings deducted
from the totalizer. Select
NO
to totalize forward
flow only and ignore reverse flow.
Flo Direction
Select
On
to enable flow direction measurement.
Select
Off
to disable flow direction measurement.
Select
Invert
to invert the sense of the flow
measurement.
Cal Constant
Scales the velocity reading. Factory value is close to
1.000 for a PSE4-A sensor.
Restore Defaults
Select
Yes
and press
to erase all user settings and return
the instrument to factory default settings.
New Password
Select any number from 0000 to 9999 and press
.
Default setting of 0000 will allow direct access to the
calibration menus. Setting of any password greater
than 0000 will require the password to be entered to
access the calibration menus.
Press
to return to
Menu Selections
.
Содержание PDFM 5.1
Страница 2: ...Page 2 Note This page has been left blank intentionally ...
Страница 6: ...Page 6 PDFM 5 1 Portable Doppler Flow Meter CALIBRATION MENU ...
Страница 17: ...Page 17 PDFM 5 1 Portable Doppler Flow Meter SENSOR MOUNTING COUPLING RECOMMENDATIONS ...
Страница 23: ...Page 23 PDFM 5 1 Portable Doppler Flow Meter FLOW METER DATA SHEET ...
Страница 31: ......