Example of Relay Calibration
Span (Max Level) = 184 inches
RON
in inches
=
150
ROFF
in inches
=
30
DAMPING
Normal Setting:
DP10
Fast Response (up to ½ inch /13 mm level change per
second):
DP5
or less
Slow Response (turbulence)
DP15
or more
NOTE: Damping Time (in seconds) is the response
time to indicate a new target. Longer Damping Times also reject false targets like agitators, splashing
etc.
INTRINSIC SAFETY BARRIER
Press
Ç
to select "
Y
" (Yes) only when the instrument is equipped
with an optional Intrinsic Safety Barrier for sensor mounting in
hazardous locations. This setting compensates for the electrical
characteristics of an Intrinsic Safety Barrier.
ECHO LOSS RESPONSE
If no echoes are received for 20 seconds or longer the 4-20mA
output can be programmed to respond as follows:
ELC1
= Hold last reading
ELC2
= Output 4mA
ELC3
= Output 20mA
When one valid echo is received the LIT25 will resume normal
operation.
Page 11
LIT25 Level Indicating Transmitter
Manual Series 3.5.1
SPAN
150"
30"
20mA = 16"
4mA = 200"
R ON
R OFF
SENSOR
È
Ç
Ç
DP11
DP12
20 SECS.
MAXIMUM
È
Ç
Ç
ELC2
ELC3
Ç
Ç
Y