Page 4
Multi-Sensor Area Velocity Flow Meter
CONNECTIONS
POWER INPUT: 100 to 240 VAC 50/60Hz. No adjustments are necessary for voltages within this
range. Connect L (Live) N (Neutral) and AC Ground.
Optional DC: 9-32 VDC. Connect to + and - terminals.
Optional Thermostat and Heater modules are available rated for 115 VAC or 230 VAC.
IMPORTANT NOTE: To comply with CSA/UL electrical safety standards, AC power input and relay
connection wires must have conduit entry to the instrument enclosure. Installation requires a switch,
overcurrent fuse or circuit breaker in the building (in close proximity to the equipment) that is marked as
the disconnect switch.
Risk of electric shock. Loosen cover screw to access connections. Only qualified personnel
should access connections.
Note: Use of instrumentation over 40°C ambient requires special field wiring.
Note: User replaceable fuse is 2 Amp 250V (T2AL250V).
FUNCTION TEST
Connect AVMS 5.1 to AVQZ3 using interface cable.
Connect sensor to the sensor A and D terminals as shown on page 6,
then apply power. Allow 30 seconds for the AVMS 5.1 to initialize.
A. Place the QZ02L sensor (flat to the bottom) in a bucket of
water about 6” deep and select
Level
mode (from
UNITS/MODE
menu) to see a level reading.
B. Select
Velocity
mode and stir the water to see velocity
reading(s).
!
Содержание AVMS 5.1