H59.0.21.6C-11
page 2 of 8
Contents
1
Intended use ..................................................................................................... 2
2
General note ..................................................................................................... 2
3
Disposal notes .................................................................................................. 2
4
Safety instructions ............................................................................................ 3
5
Operating and maintenance advice ................................................................. 4
6
Operation .......................................................................................................... 4
7
Oxygen measuring ........................................................................................... 5
8
Calibration ........................................................................................................ 5
9
Sensor rating .................................................................................................... 6
10
Replace sensor element................................................................................... 6
11
Errors and system messages ........................................................................... 6
12
Display when switching-on the device ............................................................. 6
13
Configuration of the device .............................................................................. 7
14
Specification ..................................................................................................... 8
1
Intended use
This device measures the oxygen concentration in gas mixtures and air. The actual
measurement takes place at the opening of the sensor. Due to the design of the
sensor, the device has to be calibrated at regular intervals (at fresh air =20.95%
oxygen) to get accurate measuring values. If the sensor is used up, this will be
detected at calibration and the sensor element has to be replaced before the next
measurement.
2
General note
Read this document carefully and get used to the handling of the device before you
use it. Keep this paper ready to hand in order to look it up if a question turns up.
3
Disposal notes
Dispense exhausted batteries at destined gathering places.
This device and the sensor must not be disposed as ‘residual waste’.
According to the ElektroG
(law for bringing into market, the return and the
environmentally friendly disposal of electronic equipment)
we accept the
return of this device and/or the sensor, please send it directly to us
(adequately stamped). We will dispose it appropriately and
environmentally friendly.