Sensor rating 

The rating of the sensor state is calculated and saved after each successful 
calibration.  

Watch sensor rating:

 Press shortly the 

 

key, the display shows for a short 

time i.e. 

 = 100%. 

The rating is displayed in 10 percent steps: 100% = optimal sensor status. Lower 
values are indicating that the sensor life time will soon be reached. (50% doesn’t 
mean 50% of the life time, but 50% of the reference signal!). But also a erroneous 
pressure entry may cause low sensor ratings. 

10  Replace sensor element 

Unplug the sensor, unscrew the white protection housing and take the sensor 
element. Take the rubber disk off and put it on the new sensor. Put the new sensor 
element into the housing, screw it down and plug it. 
Then start the calibration. 

11  Errors and system messages 

no  

display 

The battery is used up and has to be exchanged, or the device is 
defective 

 

The battery is used up, for a short period further measurements are 
possible 

 

The battery is used up and has to be replaced, measurements are no 
more possible  

 

Calibration error: sensor voltage is too low (sensor used up or wrong 
chosen air pressure) 

 

Calibration error: sensor voltage is too high (wrong chosen air 
pressure or sensor defective (membrane crack, etc.) 

 

Calibration error: signal not stable 
 

 

Measuring error: range has been exceeded 
 

 

Measuring error: measuring value has fallen below permitted range 
 

 

System error: the device is defective or considerably out of the 
allowed ambient temperature range 

 

12  Display when switching-on the device 

 

Start and display test 

 

If switch-

off delay “P.oF” (power off) is active, this is signalled at the 

turn-on procedure.  

 

 

Содержание GOX 100

Страница 1: ...H Messtechnik GmbH Standort Greisinger Hans Sachs Stra e 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Oxygen Measuring Device WEEE Reg Nr DE 93889386 as of V1 3 Operatin...

Страница 2: ...t fresh air 20 95 oxygen to get accurate measuring values If the sensor is used up this will be detected at calibration and the sensor element has to be replaced before the next measurement 2 General...

Страница 3: ...y to avoid re starting Operator safety may be a risk if there is visible damage to the device or the device is not working as specified the device has been stored under unsuitable conditions for a lon...

Страница 4: ...d Operation elements 1 display of the oxygen concentration in 2 BAT warning battery is used up and has to be replaced measuring is valid On Off switch Press shortly change between display of actual va...

Страница 5: ...rong air flows can produce an over pressure at the sensor possible source of measuring calibration errors The nominal life time can be shortened by Wrong storage operation temperature Permanent use wi...

Страница 6: ...it down and plug it Then start the calibration 11 Errors and system messages no display The battery is used up and has to be exchanged or the device is defective The battery is used up for a short pe...

Страница 7: ...ing 980 mbar Enter the absolute air pressure or pressure of the gas that is measured Its value depends on your altitude see the table below and the present weather 5 Press the key the current value ap...

Страница 8: ...ax Memory min and max measured value are stored Nominal temp 25 C Ambient temp 0 to 45 C sensor 20 to 50 C device Relative humidity 0 to 95 RH not condensing Storage temperature 15 to 60 C sensor 20 t...

Отзывы: