H83.0.01.6C-6
Operating manual GMH 5450
page 12 of 20
_____________________________________________________
_____________________________________________________________________________
Menu
Parameter
Value
Description
or
Set Output: Configuration of universal output
oFF
Interface and analog output off -> minimal power consumption
SEr:
Serial interface activated
dAC:
Analog output activated
01,11..91
Base address for serial interface communication
0.0000 µS/cm
.. 1000 mS/cm
Measuring value that should correspond to output 0 V
e.g. for 0.0000 µS/cm
0.0000 µS/cm
.. 1000 mS/cm
Measuring value that should correspond to output 1 V
e.g. for 100.0 mS/cm
Set Corr: Measurement correction
**
Zero point adjustment / offset of temperature measurement
**
oFF
No zero point adjustment for temperature measurement
-
5.0 … 5.0°C
Offset of temperature measurement in °C
Slope adjustment of temperature measurement
**
oFF
No slope adjustment for temperature measurement
-
5.00 … 5.00
Slope correction of temperature measurement in [%]
Set Alarm: Configuration of alarm function
On / No.So
Measuring channel cond/rES/TDS/SAL: alarm on with buzzer / without
buzzer
OFF
No alarm function for measuring channel cond/rES/TDS/SAL
0.0000 µS/cm
.. 1000 mS/cm
Min-alarm limit for cond/rES/TDS/SAL (not if AL. 1. oFF)
0.0000 µS/cm
.. 1000 mS/cm
Max-alarm limit for cond/rES/TDS/SAL (not if AL. 1. oFF)
On / No.So
Temperature measurement: alarm on with buzzer / without buzzer
OFF
No alarm function for temperature measurement
-5.0 ..+100.0 °C
Min-alarm limit for temperature (not if AL. 2. oFF)
-5.0 ..+100.0 °C
Max-alarm limit for limit temperature (not if AL. 2. oFF)
Set Logger: Configuration of logger function
**
Selection of logger function
*
oFF
No logger activated
Stor
Store
: Manual recording
CYCL
Cyclic
: Cyclic logger
0:01... 60:00
Cycle time in [minutes:seconds] (for cyclic logger)
**
Set Clock: Einstellen der Echtzeituhr
HH:MM
Clock
: set time hours:minutes
YYYY
Year
: set year
TT.MM
Date
: set date day.month
rEAd CAL: Read calibration data:
see chapter 12.2
“Calibration storage (rEAd CAL)”
(*)
If logger memory contains data sets parameters marked with (*) cannot be called. You
have to clear memory to change these parameters!
(**)
If logger is running parameters marked with (**) cannot be called.