GREISINGER GMH 5450 Скачать руководство пользователя страница 11

H83.0.01.6C-6 

 Operating manual GMH 5450 

page 11 of 20 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

 

REF.S: Choice of standard reference solutions for automatic adjustment/cal. 

 

 

1413 µS/cm 

Reference solution 0.01 M KCL 

2760 µS/cm 

0.02 M KCL 

12.88 mS/cm 

0.1 M KCL 

50 mS/cm 

Sea-water reference solution KCL 

111.8 mS/cm 

1 M KCL 

 

Unit t: Selection of temperature unit

 

 

 

°C 

All temperature values in degree Celsius  

°F 

All temperature values in degree Fahrenheit 

 

Temperature compensation (not for InP = SAL)

 

 

 

oFF 

No temperature compensation of conductivity measurement 

nLF 

Non-linear function for natural waters according to EN 27888  
(ISO 7888), ground, surface and drinking water 

NaCl 

Compensation for weak NaCl-solutions (pure and ultrapure water)  

Lin  

Linear temperature compensation 

 

Compensation coefficient (only if t.Cor = Lin) 

 

 

0.300  3.000 

Temperature compensation coefficient in %/K 

 

Reference temperature of temperature compensation (only if t.Cor = Lin or nLF) 

 

 

25 °C / 77 °F 

Reference temperature 25 °C / 77 °F 

20 °C / 68 °F 

Reference temperature 20 °C / 68 °F 

 

Adjustment/Calibration: Adjustment reminder period (factory setting: 180) 

 

 

1 …730

 

Adjustment reminder period (in days) 

oFF

 

No adjustment reminder  

 

Auto Hold: Automatic measuring value identification (only if Logger = oFF)

 

 

 

on 

Auto measuring value identification (only if Logger = oFF) Auto Hold  

oFF 

Standard hold function on keypress (only if Logger = oFF)  

 

Auto Power-Off : Selection of power-off delay 

 

 

 

1...120 

Power-off delay in minutes.  
Device will be automatically switched off as soon as this time has 
elapsed if no key is pressed/no interface communication takes place. 

 

oFF 

Automatic power-off function deactivated (continuous operation)  

 

Background illumination

 

 

 

oFF:  

Illumination deactivated 

5 … 120 

Turn off illumination after 5… 120s (factory settings: 5 s) 

 

on: 

Illumination always on 

 

 

 

Содержание GMH 5450

Страница 1: ...9402 9383 0 49 0 9402 9383 33 info greisinger de as of version 1 9 GMH 5450 WEEE Reg Nr DE 93889386 Operating manual Conductivity measuring device water proof with data logger Please carefully read th...

Страница 2: ...UREMENT 8 6 7 ELECTRODES MEASURING CELLS 9 6 7 1 Assignment bayonet connector 9 6 7 2 Design and selection 9 6 8 TEMPERATURE COMPENSATION 9 6 8 1 Temperature compensation nLF according to EN 27888 9 6...

Страница 3: ...pole bayonet socket Generally a suitable temperature sensor is included to the electrode The measured temperature is used for the automatic temperature compensation e g Lin or nlF and is additionally...

Страница 4: ...ial damage 5 This device must not be used at potentially explosive areas The usage of this device at potentially explosive areas increases danger of deflagration explosion or fire due to sparking 3 Pr...

Страница 5: ...re compensation 7 K 1 cm additional configuration units 8 logg arrow logger is ready arrow flashing automatic recording Logg CYCL is active 4 2 Pushbuttons On off key backlight press shortly activate...

Страница 6: ...ndling Pull at label open in order to swing open the pop up clip Pull at label open again to swing open the pop up clip further Pop up clip closed Pop up clip at position 90 Pop up clip at position 18...

Страница 7: ...cell constant K Application 0 01 0 004000 0 015000 1 cm Ultra pure water electrodes with K 0 01 0 1 0 04000 0 15000 1 cm Ultra pure water electrodes with K 0 1 1 0 4000 1 5000 1 cm Standard electrode...

Страница 8: ...NaCl KCl etc 0 50 Natural waters surface waters drinking water 0 65 0 70 e g salt concentration of aqueous fertilizer solutions Attention This are only approximate values good for estimations but no...

Страница 9: ...for used in petrol diesel etc with low conductivities 1000 S cm LF 210 6 8 Temperature compensation The conductivity of aqueous solutions depends on its temperature The temperature dependency is stron...

Страница 10: ...ed for more than 2 minutes the configuration will be aborted All changes will be discarded Menu Parameter Value Description or rEAd Logg Read manual recordings see chapter 8 1 Manual recording Func St...

Страница 11: ...ure compensation Compensation coefficient only if t Cor Lin 0 300 3 000 Temperature compensation coefficient in K Reference temperature of temperature compensation only if t Cor Lin or nLF 25 C 77 F R...

Страница 12: ...in Set Alarm Configuration of alarm function On No So Measuring channel cond rES TDS SAL alarm on with buzzer without buzzer OFF No alarm function for measuring channel cond rES TDS SAL 0 0000 S cm 1...

Страница 13: ...ly used for the operation of the logger functions 8 1 Manual recording Func Stor a Save measurements manually Up to 1000 measurements can be saved if logger function Func store is selected see Configu...

Страница 14: ...rage process is signaled by shortly displaying St XXXXX where XXXXX is the number of the saved data set If the logger memory is full the recording stops automatically and the display shows b Stop reco...

Страница 15: ...ur own software we offer a GMH3000 development package including a universally applicable Windows functions library GMH3x32e DLL with documentation can be used by all established programming languages...

Страница 16: ...re known by the device and are compensated automatically Press Cal key for 2 s start of calibration actual value e g 1823 S cm value of solution e g 1413 S cm and CAL with rotary symbol and CAL with r...

Страница 17: ...ut the interval after which the device reminds you to recalibrate The interval times should be chosen according to the application and the stability of the electrode CAL flashes on the display as soon...

Страница 18: ...tings can only be checked and if necessary corrected by the manufacturer A calibration protocol is enclosed to the device ex works This documents the precision reached by the production process 16 Rep...

Страница 19: ...evice CAL CAL flashing in upper display Either preset calibration interval has expired or last calibration is not valid Device has to be calibrated NO AVTO LO66 RAN6 Logger could not be started Auto r...

Страница 20: ...ext supply 4 pole bayonet socket for serial interface and supply USB Adapter USB 5100 Analog output 0 1V adjustable Display 4 digit 7 segment illuminated white Add functions Min max hold function Adju...

Отзывы: