GMH3630E.08
H36.0.02.6C-08 page 4 of 10
'Resolution': Selection of Display Resolution of Oxygen Measurements
Hi
:
high resolution, measuring ranges limited
Lo:
low resolution, maximum measuring ranges
Auto
: automatic selection of resolutions, maximum measuring ranges
2
5
max
min
2
Logg
AL
mg/l
ok
mmHg
hPa
SAL
hPa
%O
CAL
'Salinity': Selection of Salinity Correction [‰]
0 .. 70.0‰
Oxygen values will be corrected
off
:
Salinity correction deactivated (=0.0°)
2
5
max
min
2
Logg
AL
mg/l
ok
mmHg
hPa
SAL
hPa
%O
CAL
'Power.off': Selection of Power-Off Delay
2
5
max
min
2
Logg
AL
mg/l
ok
mmHg
hPa
SAL
hPa
%O
CAL
1 .. 120
: Power-off delay in minutes. Device will be automatically switched off as
soon as this time has elapsed if no key is pressed/no interface
communication takes place via the serial interface.
off
:
automatic power-off function deactivated (continuous operation, e.g. in
case of mains operation)
2
5
max
min
2
Logg
AL
mg/l
ok
mmHg
hPa
SAL
hPa
%O
CAL
'Unit P O2': Selection of Unit for Partial Oxygen Pressure hPa / mmHg
hPa:
Partial oxygen pressure in hPascal
mmHg:
Partial oxygen pressure in mm mercury column
'Unit t': Selection of Temperature Unit °C /°F
°C
:
All temperature values in degrees Celsius
°F
:
All temperature values in degrees Fahrenheit
2
5
max
min
°C
2
Logg
AL
mg/l
ok
mmHg
hPa
SAL
hPa
%O
CAL
''Offset': Zero Point Displacement When Temperature is Measured
-3.0°C .. 3.0°C
The zero point of the measurement will be displaced by the value set
or
to compensate for sensor and measuring device deviations.
-5.4°F .. 5.4°F
:
off
:
Zero displacement not activated. (=0.0°, factory setting)
temperature displayed = temperature measured - offset
2
5
max
min
2
5
max
min
Air:
Electrode calibration in air (at 100% relative humidity)
(factory setting)
AqUA:
Electrode calibration in water (oxygen saturation 100%)
'CAL': Calibration in Air or Water
SAL ok CAL
AL Logg
hPa
mg/l
%O
2
hPa
mmHg
'Lcd.2': Measuring Value Displayed in The Secondary Display
t, °C:
Permanent temperature display
P.Abs, hPa:
Permanent display of absolute atmospheric pressure
both, °C, hPa:
Both the temperature and the abs. atmospheric pressure will be
displayed (display changing at intervals)
2
5
max
min
°C
2
Logg
AL
mg/l
ok
mmHg
hPa
SAL
hPa
%O
CAL
°C
2
Logg
AL
mg/l
ok
mmHg
hPa
SAL
hPa
%O
CAL
Configuration
For configuration of the device press "Set"-key (key 4) for 2 seconds.
Choose between the individual values that can be set by pressing the "Set"-key (key 4) again.
The individual values are changed by pressing the keys "
5
" (key 2) or "
6
" (key 5).
Use key "Store" (key 6) to leave configuration and to store settings.
2
5
max
min
1-Pt:
standard 1-point calibration (factory setting)
2-Pt:
2-point calibration with the calibration equipment GKS3600, for highly
accurate measings in the upper measuring range, too (>25 mg/l)
'CAL': Choice of Calibration Mode
2
Logg
AL
mg/l
ok
mmHg
hPa
SAL
hPa
%O
CAL