9. Offset- and slope-adjustment

The offset and slope-adjustment function can be used for compensating the tolerance of the used probes.

Please note:

   The settings of the offset- / slope-adjustment will be cancelled, when no button was pressed

for more than 60 sec. Changes you may have made already won‘t be saved and will be lost!

Hint:

The buttons 2 and 3 are featured with a ‘roll-function‘. When pressing the button once the value will be raised
(button 2) by one or lowered (button 3) by one. When holding the button pressed for longer than 1 sec. the
value starts counting up or down, the counting speed will be raised after a short period of time. 
The device also features a ‘overflow-function‘, when reaching the upper limit the device switches to the lower
limit, vice versa.

- Turn on the device and wait after it finished its built-in segment test.

- Press button 3 > 2 sec.

The device will be displaying “

OFS.1

“ (offset of input 1).

- Use button 2 and button 3 for setting the desired zero point offset-value.

The input of the offset will be in digits.
The value that had been set will be subtracted from the measured value of input 1.

- Press button 1 to validate your selection. The display shows “

OFS.1

“ again.

- When pressing button 1 again, the device will be displaying “

SCA.1

“. (scale 1 = slope of input 1)

- Use button 2 and button 3 to select the desired slope-adjustment.

The slope adjustment will be entered in %. The value displayed can be calculated like this:  

Displayed value 1 = (measured value input 1 – zero point offset 1) * (1 + slope adjustment 1 [% / 100]).

Example: The setting is 2.00 => the slope has risen 2.00% => slope = 102%. 

         

When measuring a value of 1000 (without slope-adjustment) the device would display 1020 (with slope
adjustment of 102%)

- Press button 1 to validate the selection of the slope-adjustment. The display shows “

SCA.1

“ again.

- When pressing button 1 again, the device will be displaying “

OFS.2

“. (offset of input 2)

- Use button 2 and button 3 for setting the desired zero point offset-value.

The input of the offset will be in digits.
The value that had been set will be subtracted from the measured value of input 2.

- Press button 1 to validate your selection. The display shows “

OFS.2

“ again.

- When pressing button 1 again, the device will be displaying “

SCA.2

“. (scale 2 = slope of input 2)

- Use button 2 and button 3 to select the desired slope-adjustment.

The slope adjustment will be entered in %. The value displayed can be calculated like this:  

Displayed value 2 = (measured value input 2 – zero point offset 2) * (1 + slope adjustment 2 [% / 100]).

Example: The setting is 2.00 => the slope has risen 2.00% => slope = 102%. 

When measuring a value of 1000 (without slope-adjustment) the device would display 1020 (with slope
adjustment of 102%)

- Press button 1 to validate the selection of the slope-adjustment. The display shows “

SCA.2

“ again.

Now you finished configuring the device’s offset and slope adjustment. Press button 1 to switch over to dis-
play the measuring value.

Example for offset- and slope-adjustment:

Example: Connecting a temperature probe (offset deviation depending on using long wires between probe and device)
The device displays the following values (without offset- or slope-adjustment): at 0°C = 2°C, at 100°C = 102°C
Therefore you calculated: zero point: 2
 

slope: 

102 – 2 = 100 (=>deviation = 0)

You have to set:

offset =

2

(= zero point deviation)

 scale 

0.00

Button 1 Button 2 Button 3

Содержание GIR 230 NTC / DIF

Страница 1: ...E33 0 31 6C 02 Manual for connection and operation of the GIR230 Pt1000 DIF and GIR230 NTC DIF Page 1 of 14 Manual for connection and operation of GIR 230 Pt1000 DIF GIR 230 NTC DIF Version 1 3...

Страница 2: ...relay outputs 5 3 4 2 Connection of output 3 NPN output switching to ground 5 4 CONFIGURATION OF THE DEVICE 6 4 1 Selecting an input signal type 6 4 2 Selection of the output function 7 5 SWITCHPOINT...

Страница 3: ...he device must be switched off and must be marked against re starting in case of obvious malfunctions of the device which are e g visible damage the device is not working as prescribed after storage o...

Страница 4: ...factory the GIR230 has been subjected to various inspection tests and is completely cali brated Before the GIR230 can be used it has to be configured for the customer s application Hint In order to av...

Страница 5: ...switching outputs properly Please take care that you must not exceed the limits of the voltage and of the maximum current of the switching outputs not even for a short period of time Please take extr...

Страница 6: ...the device switches to the lower limit vice versa 4 1 Selecting an input signal type Turn the device on and wait until it completed its built in segment test Press button 2 for 2 sec The display will...

Страница 7: ...you have to configure the preferred states and delay time of the outputs and the switch and alarm points Depending on your output function setting it may be possible that one or more settings describ...

Страница 8: ...ss button 1 when not already done The device will be displaying 1 on turn on point of output 1 Use button 2 and button 3 to set the desired value the device s output 1 should be turning on Press butto...

Страница 9: ...points for the second switching function 2 on 2 off Configuring these settings like the first switching function After that the alarm outputs have to be configured When pressing button 1 the device w...

Страница 10: ...stays above 50 C for the entered delay time or after it had been falling below 15 C and stays below 15 C for the entered delay time Please note that the alarm outputs are inverted This means that the...

Страница 11: ...1 slope adjustment 1 100 Example The setting is 2 00 the slope has risen 2 00 slope 102 When measuring a value of 1000 without slope adjustment the device would display 1020 with slope adjustment of 1...

Страница 12: ...s check probe and device configuration Err 7 System error The device features an integrated self diagnostic function which checks essential parts of the de vice permanently When detecting a failure er...

Страница 13: ...ty 5A 230VAC resistive load NPN output NPN Open Collector breaking capacity 30mA max 28VDC Output functions 2 point 3 point 2 point with alarm 3 point with alarm min max alarm Switching points arbitra...

Страница 14: ...E33 0 31 6C 02 Manual for connection and operation of the GIR230 Pt1000 DIF and GIR230 NTC DIF Page 14 of 14...

Отзывы: