GIR1002 FR / GIA1000 FR       Connecting and Operating Manual         Version: 4.03-3       page: 15

Examples:

Data format 1:

Data format 1 corresponds to the general data format. It is only the range of values that is limited to the
display range of -1999 to +9999 equalising F831H to 270FH.
The decimal point will not be considered. The decimal point position can be called up separately.

Data format 2:

This format is used to call up and reset the system state. Data bytes D1 to D4 contain information
regarding the alarm and fault state of the GIR1002 FR.
The assignment is as follows:

Data byte D2:

Bit 0 = 1:

Fault FE1 active

Bit 1 = 1:

Fault FE2 active

Bit 2 = 1

Fault FE3 active

Bit 3 = 1

Fault FE4 active

Data byte D4:

Bit 0 = 1

max. alarm

Bit 1 = 1

min. alarm

Bit 3 = 1

alarm

Data format 3:

This format is used to call up and program the input level.
The assignment is as follows:

0 :

2Vss

1 :

300mVss

2 :

30mVss

3 :

TTL

Data format 4:

This format is used to call up and/or program the position of the decimal point.
The assignment is as follows:

0 :

no decimal point

1 :

decimal point after 2 digits (10)

2 :

decimal point after 3 digits (100)

3 :

decimal point after 4 digits (1000)

Data format 5:

This format is used to call up and program the filter.
The assignment is as follows:

0 :

filter off (FILt = 0)

1 :

filter 1 on (FILt = 1)

2 :

filter 2 on (FILt = 2)

3 :

filter 1 and filter 2 on (FILt = 3)

Data format 6:

Dieses This format is used to program the device address. Permissible device addresses range from 0 to
15.
Data format 6 corresponds to the general data format. It is only the range of values that is limited to the
permissible device addresses 0 to 15, respectively 0000H to 000FH.

Value

ASCII-character

ASCII-code

dezimal

hex

D1

D2

D3

D4

D1

D2

D3

D4

0

0000H

0

0

0

0

30H

30H

30H

30H

-1

FFFFH

F

F

F

F

46H

46H

46H

46H

-1999

F831H

F

8

3

1

46H

38H

33H

31H

9999

270FH

2

7

0

F

32H

37H

30H

46H

Содержание GIA1000 FR

Страница 1: ...rating Manual GIR1002 FR GIA1000 FR Version 4 03 This connecting and operating manual may be subject to technical alternations ON SP SET min OFF Hy 1 2 max GIR GIA ALARM Scale Analog CAL SET 1 2 3 4 I...

Страница 2: ...ge 7 Max value reset page 7 Display of min value page 7 Min value reset page 7 Display setting of making point switching point for switching output 1 or 2 page 7 Display setting of breaking point hyst...

Страница 3: ...l supply the GIR1002 FR factory tested and completely calibrated Before you can actually use it make sure to configurate your GIR1002 FR for your special application Please also refer to the chapter C...

Страница 4: ...connected to a suitable RC element for example RC220 Electric connection and commissioning of the device must be carried out by trained and skilled personnel Wrong connection may lead to the destructi...

Страница 5: ...min selectable Interface RS485 Power supply 230VAC or as specified on device housing Power consumption 5W 230VAC 3W 24VDC 1 5W 12VDC Relay switching power 10A 250V AC ohmic load use RC element when sw...

Страница 6: ...setting of making point button 1 4 1 2 3 7 display setting of switching point button 1 4 4 5 6 8 display setting of breaking point button 1 4 1 2 3 7 display setting of hysteresis button 1 4 4 5 6 8 S...

Страница 7: ...in case of alarm LED min LEDflashing Min value reset To be called up by pressing button 3 for approx 4s Display please refer to display of min value Display setting of makingpoint switching point for...

Страница 8: ...briefly no more than 1 s When pressing these buttons for a longer time over 1 s the values will start scrolling upwards respectively downwards After scrolling of approx 150 digits the scrolling speed...

Страница 9: ...int to the left or button 3 shift decimal point to the right to set desired decimal point position acknowledge position by pressing button 4 If decimal point is positioned to the right of the lowest d...

Страница 10: ...urrent of 20mA LED max will start flashing acknowledge by pressing button 4 8 LEDs Analog min and LED CAL are illuminated The display will show the measuring value stored in the GIR1002 FR for which a...

Страница 11: ...rmissibe range This fault code indicates that the min display value of 1999 digits has fallen below Remedies FE 4 will be reset as soon as the values on the display are back within their permissible r...

Страница 12: ...This method as well as the fact that the interface operates in the half duplex mode ensure that up to 16 devices can be called via only one 2 wire line Interface configuration Type RS485 Transfer mode...

Страница 13: ...s specifying the parameter and or value to be processed For assignments please refer to table 1 of chapter functional code ASCII Code 2FH Indicates the end of a data transfer process Each data transfe...

Страница 14: ...g point Switching point 0 9 30H 39H 1 Switching output 2 Breaking point Hysteresis 0 A 30H 41H 1 Max alarm limit 0 B 30H 42H 1 Min alarm limit 0 C 30H 43H 1 Input signal 1 5 31H 35H 3 Position of deci...

Страница 15: ...l The assignment is as follows 0 2Vss 1 300mVss 2 30mVss 3 TTL Data format 4 This format is used to call up and or program the position of the decimal point The assignment is as follows 0 no decimal p...

Отзывы: