12
Accuracy
Refer to “Specifications” for operating conditions and temperature
coefficient.
Accuracy is specified as follows:
±
(a percentage of the r a fixed
amount) at 23
°
C
±
5
°
C (73.4
°
F
±
9
°
F), 0% to 80% relative humidity.
Amps
Measurement
Frequency
Range
Accuracy
Range
0.0 to 40.0 A
±
(1.9% + 1 A)
50 to 60 Hz
40.0 to 200.0 A
±
(1.9% + 0.5 A)
50 to 60 Hz
201 to 400 A
±
(1.9% + 5 A)
50 to 60 Hz
AC Voltage
Measurement
Frequency
Input
Range
Accuracy
Range
Impedance
200.0 V
±
(1.5% + 0.5 V)
50 to 500 Hz
10 M
Ω
//
600 V
±
(1.5% + 5 V)
50 to 500 Hz
100 pF max
DC Voltage
Measurement
Input
Range
Accuracy
Impedance
200.0 V
±
(1.0% + 0.2 V)
10 M
Ω
//
600 V
±
(1.0% + 2 V)
100 pF max
Resistance
Measurement
Range
Accuracy
200.0
Ω
±
(1.2% + 0.5
Ω
)
2.000 k
Ω
±
(0.7% + 0.002 k
Ω
)
20.00 k
Ω
±
(0.7% + 0.02 k
Ω
)
200.0 k
Ω
±
(0.7% + 0.2 k
Ω
)
2.000 M
Ω
±
(1.0% + 0.002 M
Ω
)
20.00 M
Ω
±
(1.9% + 0.05 M
Ω
)