Accuracy
See the Specifications section for operating conditions.
Accuracy is specified as follows:
±
(a percentage of the r a fixed amount)
at 18
°
C to 28
°
C (64
°
F to 82
°
F), 0% to 80% relative humidity.
Accuracy Table
Value
Range
Accuracy
Input Impedance
1.999 V
±
(0.8% + 0.001 V)
DC Voltage
19.99 V
±
(0.8% + 0.01 V)
1 M
Ω
199.9 V
±
(0.8% + 0.1 V)
600 V
±
(0.8% + 1 V)
AC Voltage*
199.9 V
±
(1.2% + 0.3 V)
450 k
Ω
600 V
±
(1.2% + 3 V)
1.999 mA
±
(1% + 0.001 mA)
DCA
19.99 mA
±
(1.5% + 0.01 mA)
N/A
199.9 mA
±
(1.5% + 0.1 mA)
199.9
Ω
±
(1.2% + 0.2
Ω
)
1.999 k
Ω
±
(1.0% + 0.002 k
Ω
)
Resistance
19.99 k
Ω
±
(1.0% + 0.02 k
Ω
)
N/A
199.9 k
Ω
±
(1.0% + 0.2 k
Ω
)
1.999 M
Ω
±
(1.0% + 0.002 M
Ω
)
*Frequency range: 50 Hz to 500 Hz
Diode Test
Test Current:
±
1.2 mA (approximate)
Open Circuit Voltage:
±
2.8 V (approximate)