ecocool 150R
Operating Manual
Page 38
www.grantinstruments.com
6.0 Technical specifications
6.1
Operating conditions
Ambient temperature range
5 to 30°C
Altitude above sea level
Up to 2,000m (6,500ft)
Operating environment
Indoor use only
Maximum relative humidity
80% RH up to 31°C
6.2
Electrical details
Mains supply:
220-240V @ 50/60Hz or 110-120V @ 50/60Hz
Pollution degree:
2
Installation category:
II
Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.
6.3
Performance
Specification
ecocool 150R
ecocool 150 US
Typical cooling power at an
ambient of 20°C
@ 20°C
450W
@ 0°C
350W
@ -10°C
300W
@ -20°C
200W
@ -30°C
50W
Stability (DIN 12876) @ -10°C
±0.02°C
Settable temperature range
-30°C to 150°C
Tank capacity
5.5L
Top opening
120 x 155mm
Liquid depth min/max
130/145mm
Tank depth
155mm (usable 145 leaving 10mm clear)
Weight
24kg
Max pump head pressure
530mBar
Max pump flow rate
14-22 Litre/min adjustable
Safety temperature limit
User settable cut-out above 20°C
Max current consumption
12A
19A *
Heater power
1.9kW
1.4kW
*
Requires two separate mains cords
6.4
Storage and transportation
Store and transport in original packaging. Temperature range -20°C to +60°C