Grant ecocool 100R Скачать руководство пользователя страница 31

 

 ecocool 100R 

 

Operating Manual 

Page 30 

www.grantinstruments.com 

11.0  Troubleshooting 

 

Symptom 

Possible cause 

Action required 

Display shows 



” 

Refrigerated bath is 
not detected by 
heating circulator 
 
 
Refrigerated bath has 
overheated 

Check communications lead is connected to 
control and refrigerated baths 
Check refrigerated bath is connected to 
mains power and switched on. 
 
Wait for compressor to cool, when thermostat 
will reset and compressor will switch on 
again. 
 
Remove any vent obstructions and carry out 
routine cleaning (see section 9.2) 

Display shows 



” 

Refrigerated bath is 
connected but not 
powered 

Check mains power to refrigerated bath. 

Display shows 



” 

Chiller in pre-over 
heating condition 

Switch off equipment. Check air vents are not 
restricted. Clean dust from condenser fins. 
See Section 9. 

Display shows 



” 

Over-temperature 
cut-out has operated 

Check the set temperature is correct and that 
the over-temperature cut-out temperature is 
set at least 5°C above the set temperature. 
Refer to section 5.1.2 for setting instructions. 
 
If the over-temperature cut-out temperature is 
correctly set but the equipment still shows a 
“Cut” alarm then the equipment has an 
internal fault and must be repaired before it is 
used again. 

Temperature 
does not rise 
when expected 

Set temp is lower 
than liquid temp 
  

Check that the bath set temperature is correct 
(see section 5.2.1). 
  

Temp continues 
to rise when not 
expected 

Set temp is higher 
than liquid temp 

 

Check that the bath set temperature is correct 
(see section 5.2.1). 

 

Display shows 



” 

Refrigerated bath has 
stopped cooling 
 
Refrigerated bath has 
overheated 

Check communications lead is connected to 
control and refrigerated baths. 
 
Switch off refrigerated bath and allow to cool 

Alarm light on 

Over-temperature 
cut-out has operated 
 
Low liquid level float 
switch has operated 

Check the over-temperature cut-out is set 
appropriately (see section 5.1.2). 
 
Check that the liquid level in the bath is 
adequate. 

Set temperature 
too restricted 

Liquid type set does 
not allow required set 
point 

Change to different liquid type (see section 
5.2.2). 

Содержание ecocool 100R

Страница 1: ...Combined refrigerated and heating bath circulator ecocool 100R Operating Manual EN...

Страница 2: ...t Instruments Cambridge Ltd Shepreth Cambridgeshire SG8 6GB UK Tel 44 0 1763 260 811 E mail support grantinstruments com Representative in the European Union Grant Instruments Europe B V Strawinskylaa...

Страница 3: ...15 5 1 6 Setting up and switching on 16 5 1 7 Eco mode cooling 16 5 2 Using the equipment 17 5 2 1 Setting the control temperature 17 5 2 2 Selecting liquid type 17 5 2 3 Configuring a pre set 17 5 2...

Страница 4: ...ents com 9 5 Routine safety tests 26 9 6 Disposal 27 9 7 Service 27 10 0 Optional accessories 28 10 1 Alternative pump 28 10 2 Pump connectors and other accessories 28 11 0 Troubleshooting 30 12 0 Con...

Страница 5: ...se this operating manual This operating manual will allow you to unpack set up and operate this refrigerated and heating bath circulator correctly and safely Important safety information symbols and w...

Страница 6: ...the manufacturer the protection provided by the equipment may be impaired This equipment is only for indoor use by a professional user Use only water water glycol or silicone oil with this equipment...

Страница 7: ...to ambient temperature and wait until all the ice has melted WARNING Do not damage the refrigerant circuit 4 0 Operating instructions 4 1 Unpacking instructions Standard equipment includes ecocool100R...

Страница 8: ...to low liquid level Over temperature cut out has operated Temperature probe fault The cause of the alarm will also be shown on the display The display normally shows the temperature of the liquid bei...

Страница 9: ...ain connected at all times Only use the communications port cable supplied with the system Do not connect any other cables or equipment to this communication port Mains Power cable inlet and switch Fo...

Страница 10: ...rain insert Place the non connected end of tubing in a receptacle suitable for the liquid to be drained Push the drain insert into the drain port and let the liquid drain To release the drain insert p...

Страница 11: ...supplied with the system Do not connect any other cables or equipment to this communication port Mains power outlet for heating circulator 230V equipment only Provides power for 230V heating circulat...

Страница 12: ...mains plug and the switch at the rear of the equipment are easily accessible After transport or storage in humid conditions always allow the equipment to stand for at least an hour at room temperatur...

Страница 13: ...5 15P plug suitable for North American outlets 230 V equipment Connect the heating circulator using the mains link cable by plugging it into the Mains power inlet on the heating circulator see figure...

Страница 14: ...silicone fluids at elevated temperatures Temp range Recommended liquid Comments 20 C to 50 C 50 water 50 antifreeze inhibited ethylene glycol Ethylene glycol is toxic follow the manufacturer s instruc...

Страница 15: ...ed at the top right of the heating circulator The over temperature probe independently monitors the bath temperature and switches the heater off if it goes above the cut out threshold Coarse setting o...

Страница 16: ...nection Retain the blanking plate for refitting when the pump is no longer required The equipment is supplied with insulated hose which is designed to operate at temperatures between 40 C to 100 C Sui...

Страница 17: ...g circulator has started up When working at high temperatures ensure the cables are not touching hot surfaces or air vents Switch on the equipment using the power switch on the front of the refrigerat...

Страница 18: ...on s responsibility to assess the risks associated with the choice of heat transfer liquid chosen and to put in place safe working practises and appropriate protective equipment for personnel when con...

Страница 19: ...button and rotate the dial until the display shows the correct preset 2 Press the S button to select the preset 3 Press the S button to set the bath to the preset temperature 5 2 5 Setting a countdown...

Страница 20: ...the heating circulator will sound the buzzer intermittently and show on the display The alarm can be acknowledged by pressing either the F or S buttons However the alarm is persistent and will trigge...

Страница 21: ...f the bath or vessel The calibration of the equipment is performed by setting a low temperature calibration point and a high temperature calibration point The high temperature calibration point must b...

Страница 22: ...he temperature on the reference thermometer 7 Press the S button to store the temperature Press the F button to cancel the calibration 5 2 11 Restoring factory calibration settings The restore factory...

Страница 23: ...ceed 10 of the nominal supply voltage 6 3 Performance Specification ecocool 100R ecocool 100R AUS ecocool 100R US Typical cooling power at an ambient temperature of 20 C 20 C 250W 0 C 200W 10 C 100W 2...

Страница 24: ...e grade of stainless steel It is the excellent forming characteristic that has made this grade dominant in the manufacture of laboratory and industrial water baths as well as domestic sinks and saucep...

Страница 25: ...uncontrolled the growth of these organisms can result in a range of serious problems and health risks from pathogenic bacteria The growth of algae on the surfaces will cause bio fouling which can redu...

Страница 26: ...ut can be reset correctly If the alarm lamp fails to illuminate when the value indicated on the over temperature dial is more than 10 C below the current temperature as indicated by the main display t...

Страница 27: ...replaceable fuse Replacement fuses must be 20mm x 5mm Quick blow type with a F2 5A L 250V rating Fuse replacement should only be carried out by a competent person 9 4 Replacing the mains cord This eq...

Страница 28: ...Grant Instruments for disposal 9 7 Service If service is required switch off the equipment and contact Grant Instruments or your local representative for repairs Please note all returned equipment mus...

Страница 29: ...pump inlet outlet connector fits 15mm ID T range 50 to 200 C M M16 Stainless steel pump inlet outlet connector M16 x 1 male T range 50 to 200 C M SR4 Metal pump inlet outlet connector dual seal super...

Страница 30: ...umentation for the ecocool 100R PQ ecocool 100R PQ documentation for the ecocool 100R A full listing of equipment accessories and options is available in the Grant Scientific Reference Catalogue a cop...

Страница 31: ...e is set at least 5 C above the set temperature Refer to section 5 1 2 for setting instructions If the over temperature cut out temperature is correctly set but the equipment still shows a Cut alarm t...

Страница 32: ...rature probe Have a competent person check the probe for an open circuit fault or contact Grant Display shows Faulty internal temperature probe Have a competent person check the probe for a short circ...

Страница 33: ...tton to return to the home screen Display shows manufacturer factory setting menu There are no user configuration options in this menu Press the F button to return to the home screen For any other err...

Страница 34: ...tances hazardous to health RoHS This product complies with the requirements of the UK Restriction of the Use of Certain Substances in Electrical and Electronic equipment Regulations 2012 and the EU Ro...

Страница 35: ...ecocool 100R Operating Manual Page 34 www grantinstruments com Notes...

Страница 36: ...ts Europe B V Shepreth Strawinskylaan 411 Cambridgeshire WTC Tower A 4th Floor SG8 6GB 1077 XX AMSTERDAM UK THE NETHERLANDS Tel 44 0 1763 260811 grant eu grantinstruments com salesdesk grantinstrument...

Отзывы: