6.0 Technical specifications
6.1
Operating conditions
Ambient temperature range
5 to 40°C
Altitude above sea level
Up to 2,000m (6,500ft)
Operating environment
Indoor use only
Maximum relative humidity
80% RH up to 31°C decreasing to 50% RH at 40°C
6.2
Electrical details
Mains supply:
220-240V @ 50 or 60Hz* or 110-120V @ 50/60Hz
Pollution degree:
2
Installation category:
II
*220-240V heating circulators can only operate at a single mains frequency. Check the
rating plate of the heating circulator.
Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.
6.3
Performance
Specification
ecocool 100R
ecocool 100R AUS
ecocool 100R US
Typical cooling power at an
ambient temperature of 20°C
@ 20°C
250W
@ 0°C
200W
@ -10°C
100W
@ -20°C
50W
Stability (DIN 12876) @ -10°C
±0.05°C
Settable temperature range
-20°C to 100°C
Tank capacity
5L
Top opening
120 x 155mm
Tank depth
150mm (usable 140 leaving 10mm clear)
Liquid depth min/max
85/145mm
Weight
29kg
Max pump head pressure
250mBar
Max pump flow rate
17L/min
Safety high temperature limit
Adjustable cut-out above 20°C
Max current consumption (both equipment
excluding external pump)
7A
18A*
Heater power
1.3kW
1.4kW
*
Combined rating of both the refrigeration bath and heating circulator.
6.4
Storage and transportation
Store and transport in original packaging. Temperature range -20°C to +60°C