![Grant ecocool 100R Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/grant/ecocool-100r/ecocool-100r_operating-manual_2256787019.webp)
ecocool 100R
Operating Manual
Page 18
www.grantinstruments.com
Be aware that once a preset has been saved, it will automatically operate once the steps
below have been completed.
1. Press the
F
button on the heating circulator and rotate the dial until the display shows
“
”.
The display will alternate between the preset number and its temperature.
2. Press the
S
button to select the preset.
3. Use the dial to set the desired preset temperature.
If no key is pressed for 10 seconds then the display will revert back to showing the
bath temperature and the preset temperature will remain at its original value.
4. Press the
S
button to save the preset temperature.
The preset will automatically start as soon as the value is saved.
Preset temperatures and set temperatures are limited to between -20°C and 100°C. The
factory preset settings are shown in the following table.
Preset
Set temperature range
t-1
37°C
t-2
56°C
t-3
72°C
5.2.4 Running a bath preset
1. Press the
F
button and rotate the dial until the display shows the correct preset.
2. Press the
S
button to select the preset.
3. Press the
S
button to set the bath to the preset temperature.
5.2.5 Setting a countdown timer
The heating circulator includes a countdown timer in minutes which triggers an audible
alarm on completion. The timer can be set between 1 minute and 6000 minutes (100
hours).
1. Press the
F
button twice
– the display will show “
”.
2. Press the
S
button to select.
3. Use the main dia
l to select “
” and press the
S
button to select. The display flashes
with time (in minutes) indicated. Rotate navigator control to set time required.
4. Press the
S
button to save and return to liquid temperature display with timer
countdown started.
When the countdown clock is set the display alternates between bath temperature and
time remaining in minutes.
When the countdown timer reaches zero the audible alarm will sound and the display will
alternate between the water temperature and
“
”. The alarm can be accepted by
pressing the either the
F
or
S
buttons.
Note: the equipment will continue to control at the set temperature after the timer
reaches zero
– the heater will not switch off.