Glossary of Terms
14
312776K
Overdose (A, B, C) Alarm
- when either the resin (A),
or catalyst (B), or reducer (C) component dispenses too
much material and the system cannot compensate for
the additional material.
Potlife Time
- the amount of time before a material
becomes unsprayable.
Potlife Volume
- the amount of material that is required
to move through the mix manifold, hose and applicator
before the potlife timer is reset.
Purge
- when all mixed material is flushed from the sys-
tem.
Purge Drive
- The voltage drive during the Purge
sequence, maximum of 3300 mV. The response curve
of the V/P regulator is not linear, so it may be necessary
to test the response using Manual Override mode.
Purge Time
- the amount of time required to flush all
mixed material from the system.
Purge Volume Alarm
- E-11 Alarm occurs if minimum
flush volume is not achieved.
Ratio Tolerance
- the settable percent of acceptable
variance that the system will allow before a ratio alarm
occurs.
Sequential Color Change
- the process when a color
change is initiated and the system automatically flushes
the old color and loads a new color.
Sequential Dosing
- Components A and B dispense
sequentially in the necessary volumes to attain the mix
ratio.
Solvent/3rd Purge Valve Chop Time
- duration of each
activation of the solvent or 3rd purge valve during a
chop sequence. User settable from 0.0-99.9 seconds.
Solvent Fill
- the time required to fill the mixed material
line with solvent.
Solvent Push
- enables the user to save some mixed
material by pushing it out to the gun with solvent.
Requires an accessory solvent meter.
Standby
- refers to the status of the system.
System Idle
- This warning occurs if the ProMix is set to
Mix, and 2 minutes have elapsed since the system
received a flow meter pulse.
Third Purge Valve
- refers to the use of three purge
valves used to flush some waterborne materials. The
valves are used to flush with water, air and solvent.
V/P
- refers to the voltage to pressure device in the flow
control module.
Valve Holdoff Maximum
- The maximum amount of
time that flow rate learning is not allowed after a dose
valve cycles. The system may internally use a time less
than is based on the stability of the fluid meter pulse
stream.